Optimal distance of multi-plane sensor in three-dimensional electrical impedance tomography.

Comput Assist Surg (Abingdon)

a School of Electrical and Information Engineering , Tianjin University, Tianjin , China.

Published: December 2017

AI Article Synopsis

  • - Electrical impedance tomography (EIT) is a imaging technology that measures conductivity and permittivity distributions, showing promise for monitoring lung function continuously at the bedside.
  • - The technique faces challenges with 3D effects, which can distort measurements; using multiple sensor groups can help counter these issues.
  • - The study aims to identify the best spacing between electrode planes in a 3D EIT setup, employing simulations and experiments to determine optimal configurations for accurate imaging in lung function monitoring.

Article Abstract

Electrical impedance tomography (EIT) is a visual imaging technique for obtaining the conductivity and permittivity distributions in the domain of interest. As an advanced technique, EIT has the potential to be a valuable tool for continuously bedside monitoring of pulmonary function. The EIT applications in any three-dimensional (3 D) field are very limited to the 3 D effects, i.e. the distribution of electric field spreads far beyond the electrode plane. The 3 D effects can result in measurement errors and image distortion. An important way to overcome the 3 D effect is to use the multiple groups of sensors. The aim of this paper is to find the best space resolution of EIT image over various electrode planes and select an optimal plane spacing in a 3 D EIT sensor, and provide guidance for 3 D EIT electrodes placement in monitoring lung function. In simulation and experiment, several typical conductivity distribution models, such as one rod (central, midway and edge), two rods and three rods, are set at different plane spacings between the two electrode planes. A Tikhonov regularization algorithm is utilized for reconstructing the images; the relative error and the correlation coefficient are utilized for evaluating the image quality. Based on numerical simulation and experimental results, the image performance at different spacing conditions is evaluated. The results demonstrate that there exists an optimal plane spacing between the two electrode planes for 3 D EIT sensor. And then the selection of the optimal plane spacing between the electrode planes is suggested for the electrodes placement of multi-plane EIT sensor.

Download full-text PDF

Source
http://dx.doi.org/10.1080/24699322.2017.1389412DOI Listing

Publication Analysis

Top Keywords

electrode planes
16
optimal plane
12
plane spacing
12
3 d eit
12
eit sensor
12
electrical impedance
8
impedance tomography
8
eit
8
3 d effects
8
electrodes placement
8