Generating method-specific Reference Ranges - A harmonious outcome?

Pract Lab Med

Department of Clinical Biochemistry and Diagnostic Endocrinology, Mater Misericordiae University Hospital, Eccles Street, Dublin, D7, Ireland.

Published: December 2017

Objectives: When laboratory Reference Ranges (RR) do not reflect analytical methodology, result interpretation can cause misclassification of patients and inappropriate management. This can be mitigated by determining and implementing method-specific RRs, which was the main objective of this study.

Design And Methods: Serum was obtained from healthy volunteers (Male + Female, n > 120) attending hospital health-check sessions during June and July 2011. Pseudo-anonymised aliquots were stored (at - 70 °C) prior t° analysis on Abbott ARCHITECT c16000 chemistry and 2000SR immunoassay analysers. Data were stratified by gender where appropriate. Outliers were excluded statistically (Tukey method) to generate non-parametric RRs (2.5th + 97.5th percentiles). RRs were compared to those quoted by Abbott and UK Pathology Harmony (PH) where possible. For 7 selected tests, RRs were verified using a data mining approach.

Results: For chemistry tests (n = 23), Upper or Lower Reference Limits (LRL or URL) were > 20% different from Abbott ranges in 25% of tests (11% from PH ranges) but in 38% for immunoassay tests (n = 13). RRs (mmol/L) for sodium (138-144), potassium (3.8-4.9) and chloride (102-110) were considerably narrower than PH ranges (133-146, 3.5-5.0 and 95-108, respectively). The gender difference for ferritin (M: 29-441, F: 8-193 ng/mL) was more pronounced than reported by Abbott (M: 22-275, F: 5-204 ng/mL). Verification studies showed good agreement for chemistry tests (mean [SD] difference = 0.4% [1.2%]) but less so for immunoassay tests (27% [29%]), particularly for TSH (LRL).

Conclusion: Where resource permits, we advocate using method-specific RRs in preference to other sources, particularly where method bias and lack of standardisation limits RR transferability and harmonisation.

Download full-text PDF

Source
http://www.ncbi.nlm.nih.gov/pmc/articles/PMC5633846PMC
http://dx.doi.org/10.1016/j.plabm.2017.06.001DOI Listing

Publication Analysis

Top Keywords

reference ranges
8
method-specific rrs
8
tests rrs
8
chemistry tests
8
immunoassay tests
8
rrs
6
tests
6
ranges
5
generating method-specific
4
method-specific reference
4

Similar Publications

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!