Purpose: To investigate the reliability of diffusion weighted image (DWI) measurements obtained on a 0.35T MR scanner in Malawi for malaria research.

Materials And Methods: The same healthy volunteers (n=6) were scanned on a 0.35T MR scanner in Malawi and a 3T scanner in the US. Three subjects had two repeated DWI scans at 0.35T. Due to scanner constraints, only three diffusion gradient directions for DWI on 0.35T could be obtained. An apparent diffusion coefficient (ADC) map was reconstructed from the 0.35T and the result was compared to standard DWI acquisition on the 3T scanner. The mean ADC from 15 different regions and the voxel-wise coefficient of variation (CV) were calculated to investigate the intra-scanner and inter-scanner variability. Reproducibility was calculated using intra-class correlation coefficient (ICC).

Results: The 0.35T intra-scanner ADC repeatability was high for all three subjects with repeated scans (ICC>0.7). The intra-scanner correlation between repeated scans was also high (r>0.67, p< 0.01). Comparing the ADC findings from the 0.35T and 3T MRs, the high inter-scanner correlation suggested that the 0.35T ADC results were valid (ICC>0.7, r>0.5, p<0.01). Voxel-wise CV revealed a few regions with larger variation (CV>20%), which were primarily located in peripheral regions and the boundary of lateral ventricles, and likely due to partial volume effects in low field scans.

Conclusion: These findings support the validity of DWI obtained from low field MR scanners used in many low income countries.

Download full-text PDF

Source
http://dx.doi.org/10.1016/j.mri.2017.10.001DOI Listing

Publication Analysis

Top Keywords

low field
12
035t scanner
12
035t
9
measurements 035t
8
low income
8
scanner malawi
8
three subjects
8
subjects repeated
8
repeated scans
8
low
5

Similar Publications

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!