Purpose: To investigate the accuracy of the measurements provided by a new optical biometer (OA-2000, Tomey Corporation, Nagoya, Japan) for calculating the intraocular lens (IOL) power and to compare the refractive outcomes to those obtained with the IOLMaster 500 (Carl Zeiss Meditec, Jena, Germany).

Methods: In this interventional multicenter study, consecutive patients having cataract surgery were enrolled. Only the IOL model used in the largest sample of patients was selected and the eyes implanted with that IOL were subsequently analyzed. The OA-2000, an optical biometer based on swept-source optical coherence tomography (SS-OCT), was used to measure axial length and corneal power in all eyes. IOL power was calculated with the Hoffer Q, Holladay 1, and SRK/T formulas. In a subsample of eyes, the IOL power was also calculated with measurements obtained by partial coherence interferometry (IOLMaster 500). The median absolute error and percentage of eyes with a prediction error of ±0.50 diopters (D) or less were calculated.

Results: Two hundred forty-nine eyes were analyzed. Using SS-OCT, the median absolute error ranged between 0.33 (Holladay 1) and 0.35 (SRK/T) D. The rate of eyes with a prediction error of ±0.50 D or less ranged between 71.5% (Hoffer Q) and 67.1% (SRK/T). In the subsample of 87 eyes with measurements by both devices, the median absolute error was lower with the OA-2000 (Hoffer Q: P = .0377; Holladay 1: P = .0191; SRK/T: P = .0087).

Conclusions: The SS-OCT-based optical biometer investigated in the current study provides accurate measurements for IOL power calculation and seems to offer more predictable refractive results compared to the partial coherence interferometry IOLMaster. [J Refract Surg. 2017;33(10):690-695.].

Download full-text PDF

Source
http://dx.doi.org/10.3928/1081597X-20170721-05DOI Listing

Publication Analysis

Top Keywords

iol power
20
optical biometer
12
median absolute
12
absolute error
12
swept-source optical
8
optical coherence
8
coherence tomography
8
power calculation
8
iolmaster 500
8
eyes iol
8

Similar Publications

Impact of central corneal astigmatism on postoperative visual outcomes in patients undergoing trifocal intraocular lens implantation.

BMC Ophthalmol

January 2025

Department of Ophthalmology, Eye & ENT Hospital, Fudan University, 83 Fenyang Road, Xuhui District, Shanghai, 200031, China.

Background: To investigate the impact of central corneal astigmatism on postoperative visual outcomes in patients with trifocal intraocular lens (IOL) implantation.

Methods: This retrospective study included 278 eyes of 278 patients who underwent uneventful cataract surgery with implantation of the trifocal IOL (AT LISA tri 839MP). Patients were divided into two groups according to the total corneal refractive power (TCRP) in 3 mm zone centered on the corneal apex: low astigmatism group, TCRP ≤ 0.

View Article and Find Full Text PDF

Background: Extended monovision is a novel mix-and-match approach that has been recently introduced. It involves implanting an aspherical monofocal intraocular lens (IOL) for distance vision in the dominant eye, and a bifocal extended depth-of-focus (EDOF) IOL in the nondominant eye. The target refraction for the nondominant eye is - 1.

View Article and Find Full Text PDF

Background: To determine whether accounting for posterior corneal surgically induced astigmatism (SIA) would improve toric intraocular lens power calculation prediction error.

Methods: A total of 189 eyes of 148 patients undergoing routine cataract surgery were included in the study. Standard and posterior keratometry were measured pre- and postoperatively.

View Article and Find Full Text PDF

Purpose: To investigate different measures for corneal astigmatism in the context of reconstructed corneal astigmatism (recCP) as required to correct the pseudophakic eye, and to derive prediction models to map measured corneal astigmatism to recCP.

Methods: Retrospective single centre study of 509 eyes of 509 cataract patients with monofocal (MX60P) IOL. Corneal power measured with the IOLMaster 700 keratometry (IOLMK), and Galilei G4 keratometry (GK), total corneal power (TCP2), and Alpin's integrated front (CorT) and total corneal power (CorTTP).

View Article and Find Full Text PDF

Objectives: This study compared the visual outcomes and optical quality of two monofocal, aspheric intraocular lenses (IOLs; CT LUCIA 621P, Carl Zeiss Meditec; Eyhance ICB00, Johnson & Johnson Vision) by evaluating visual acuity, contrast sensitivity, and higher-order aberrations 1 month post-cataract surgery.

Methods: In this retrospective, comparative study, 120 eyes (72 patients) that underwent cataract surgery with either CT LUCIA 621P (Lucia group) or Eyhance ICB00 (Eyhance group) implantation (60 eyes/group) were retrospectively investigated. Visual acuity at various distances and defocus curves were measured 1 month postoperatively.

View Article and Find Full Text PDF

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!