Objectives: Magnetic resonance imaging (MRI) is regarded as a non-harming and non-invasive imaging modality with high tissue contrast and almost no side effects. Compared to other cross-sectional imaging modalities, MRI does not use ionising radiation. Recently, however, strong magnetic fields as applied in clinical MRI scanners have been suspected to induce DNA double-strand breaks in human lymphocytes.

Methods: In this study we investigated the impact of 3-T cardiac MRI examinations on the induction of DNA double-strand breaks in peripheral mononuclear cells by γH2AX staining and flow cytometry analysis. The study cohort consisted of 73 healthy non-smoking volunteers with 36 volunteers undergoing CMRI and 37 controls without intervention. Differences between the two cohorts were analysed by a mixed linear model with repeated measures.

Results: Both cohorts showed a significant increase in the γH2AX signal from baseline to post-procedure of 6.7 % (SD 7.18 %) and 7.8 % (SD 6.61 %), respectively. However, the difference between the two groups was not significant.

Conclusion: Based on our study, γH2AX flow cytometry shows no evidence that 3-T MRI examinations as used in cardiac scans impair DNA integrity in peripheral mononuclear cells.

Key Points: • No evidence for DNA double-strand breaks after cardiac MRI. • Prospective study underlines safe use of MRI with regard to DNA damage. • Controlled trial involving both genders investigating DNA DSBs after 3-T MRI.

Download full-text PDF

Source
http://dx.doi.org/10.1007/s00330-017-5056-9DOI Listing

Publication Analysis

Top Keywords

dna double-strand
16
double-strand breaks
16
flow cytometry
12
breaks human
8
mononuclear cells
8
magnetic resonance
8
resonance imaging
8
γh2ax flow
8
mri
8
cardiac mri
8

Similar Publications

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!