AI Article Synopsis

Article Abstract

Statement Of Problem: Little information is available on the effect of drilling speed on surrounding bone during the removal of an abutment screw fragment.

Purpose: The purpose of this in vitro study was to compare, in vitro, the peak temperature increase during the removal of fractured abutment screws from implants placed in a porcine mandible, using drilling speeds of 600 or 2000 rpm.

Material And Methods: Twenty 4.3×13-mm dental implants were placed in 10 dissected porcine mandibles: 2 implants per mandible, 1 on each side. Localized defects were created in 20 surface-treated abutment screws, which were then tightened into each implant until a reproducible fracture occurred in each screw. The fractured screws were removed with a handpiece removal kit and irrigated with room-temperature water at either 600 or 2000 rpm. The temperature rise at the implant surface was measured at 3 levels with 3 type-K thermocouples. Repeated measure ANOVA was performed with the Tukey-Kramer post hoc test for mean pair-wise comparisons (α=.05 for all tests).

Results: Mean peak temperatures were significantly higher at 2000 rpm than at 600 rpm in the mid-body (P<.001) and crestal (P=.003) regions but not in the apical (P=.225) implant locations. No significant differences in mean peak temperatures were found among the 3 locations using 600 rpm (P=.179). In the 2000-rpm group, mean peak temperature in the mid-body area was consistently higher than that in the apical (P<.001) area, and more instances of temperature rise above 56°C and 60°C were observed. In 1 implant from this group, the estimated peak temperature exceeded the bone damage threshold value (50°C for 30 seconds).

Conclusions: A drilling speed of 2000 rpm during the removal of abutment screw fragments caused overheating of the outer surface of the implant which may damage the surrounding bone; a speed of 600 rpm appears to be safe.

Download full-text PDF

Source
http://dx.doi.org/10.1016/j.prosdent.2017.06.011DOI Listing

Publication Analysis

Top Keywords

removal abutment
8
abutment screw
8
dental implants
8
abutment screws
8
600 2000
8
2000 rpm
8
heat generation
4
removal
4
generation removal
4
abutment
4

Similar Publications

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!