Background: Drawing Archimedes spirals is a popular and valid method of assessing action tremor in the upper limbs. We performed the first blinded comparison of Fahn-Tolosa-Marín (FTM) ratings and tablet measures of essential tremor to determine if a digitizing tablet is better than 0-4 ratings in detecting changes in essential tremor that exceed random variability in tremor amplitude.
Methods: The large and small spirals of FTM were drawn with each hand on two consecutive days by 14 men and four women (age 60±8.7 years [mean±SD]) with mild to severe essential tremor. The drawings were simultaneously digitized with a digitizing tablet. Tremor in each digitized drawing was computed with spectral analysis in an independent laboratory, blinded to the clinical ratings. The mean peak-to-peak tremor displacement (cm) in the four spirals and mean FTM ratings were compared statistically.
Results: Test-retest intraclass correlations (ICCs) (two-way random single measures, absolute agreement) were excellent for the FTM ratings (ICC 0.90, 95% CI 0.76-0.96) and tablet (ICC 0.97, 95% CI 0.91-0.99). Log tremor amplitude () and FTM were strongly correlated (log=αFTM + β, α≈0.6, β≈-1.27, r=0.94). The minimum detectable change for the tablet and FTM were 51% and 67% of the initial assessment.
Discussion: Digitizing tablets are much more precise than clinical ratings, but this advantage is mitigated by the natural variability in tremor. Nevertheless, the digitizing tablet is a robust method of quantifying tremor that can be used in lieu of or in combination with clinical ratings.
Download full-text PDF |
Source |
---|---|
http://www.ncbi.nlm.nih.gov/pmc/articles/PMC5618112 | PMC |
http://dx.doi.org/10.7916/D89S20H7 | DOI Listing |
Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!