  • A new molecular method for monitoring the cleanliness of endoscopes after colonoscopy was tested, showcasing quicker results than traditional culture methods.
  • This method involved sampling with special swabs and using a real-time PCR test to detect bacteria, proving effective across multiple hospitals in Italy.
  • The study demonstrated a high level of accuracy (98% sensitivity and specificity) in identifying contaminated versus properly cleaned endoscopes, potentially improving hospital hygiene protocols.

Article Abstract

Background: Reprocessing of endoscopes is key to preventing cross-infection after colonoscopy. Culture-based methods are recommended for monitoring, but alternative and rapid approaches are needed to improve surveillance and reduce turnover times. A molecular strategy based on detection of residual traces from gut microbiota was developed and tested using a multicenter survey.

Methods: A simplified sampling and DNA extraction protocol using nylon-tipped flocked swabs was optimized. A multiplex real-time polymerase chain reaction (PCR) test was developed that targeted 6 bacteria genes that were amplified in 3 mixes. The method was validated by interlaboratory tests involving 5 reference laboratories. Colonoscopy devices (n = 111) were sampled in 10 Italian hospitals. Culture-based microbiology and metagenomic tests were performed to verify PCR data.

Results: The sampling method was easily applied in all 10 endoscopy units and the optimized DNA extraction and amplification protocol was successfully performed by all of the involved laboratories. This PCR-based method allowed identification of both contaminated (n = 59) and fully reprocessed endoscopes (n = 52) with high sensibility (98%) and specificity (98%), within 3-4 hours, in contrast to the 24-72 hours needed for a classic microbiology test. Results were confirmed by next-generation sequencing and classic microbiology.

Conclusions: A novel approach for monitoring reprocessing of colonoscopy devices was developed and successfully applied in a multicenter survey. The general principle of tracing biological fluids through microflora DNA amplification was successfully applied and may represent a promising approach for hospital hygiene.

Download full-text PDF

Source
http://dx.doi.org/10.1016/j.ajic.2017.08.008DOI Listing

Publication Analysis

Top Keywords

colonoscopy devices
12
real-time polymerase
8
polymerase chain
8
chain reaction
8
dna extraction
8
potential testing
4
testing reprocessing
4
reprocessing procedures
4
procedures real-time
4
reaction multicenter
4

Similar Publications

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!