Severity: Warning
Message: file_get_contents(https://...@pubfacts.com&api_key=b8daa3ad693db53b1410957c26c9a51b4908&a=1): Failed to open stream: HTTP request failed! HTTP/1.1 429 Too Many Requests
Filename: helpers/my_audit_helper.php
Line Number: 176
Backtrace:
File: /var/www/html/application/helpers/my_audit_helper.php
Line: 176
Function: file_get_contents
File: /var/www/html/application/helpers/my_audit_helper.php
Line: 250
Function: simplexml_load_file_from_url
File: /var/www/html/application/helpers/my_audit_helper.php
Line: 3122
Function: getPubMedXML
File: /var/www/html/application/controllers/Detail.php
Line: 575
Function: pubMedSearch_Global
File: /var/www/html/application/controllers/Detail.php
Line: 489
Function: pubMedGetRelatedKeyword
File: /var/www/html/index.php
Line: 316
Function: require_once
Background And Objectives: Balloon sizing remains the main technique for determining occluder device size for atrial septal defects (ASDs). New evidence has proposed that accurate estimation of device size could be possible without using the balloon technique. Operators have predicted the amount of possible enlargement depending on their experiences. Thus, selection criteria have mostly relied on personal observations and experiences. The objective of this study was to determine the relationship between age, sex, defect size, and deployed device size based on the balloon technique.
Subjects And Methods: Sixty-six patients who underwent percutaneous ASD closure with a Cardi-O-Fix occluder between 2011 and 2012 were retrospectively evaluated. Patients whose maximum defect size and device size were available were included. Enlargement amount (EA) (device size-defect size) and enlargement ratio (ER) (EA/defect size) were calculated. The relationship between these 2 calculations and age, sex, and defect size were analyzed.
Results: EA and ER were 5.2±3.6 mm (min: 0, max: 15, median: 5) and 39.3%±31.5% (min: 0, max: 125, median: 32), respectively. EA and ER did not differ between genders (p=0.800; p=0.430). EA and ER were not correlated with maximum defect size (p=0.310; p=0.050). EA and ER showed no correlation with age (p=0.970; p=0.640). However when patients were dichomotized based on age 40, ER was significantly lower in older group (p=0.030). Unexpectedly, no difference was observed between the 2 groups in terms of EA (p=0.110). Size of deployed device had a strong correlation with defect size measured with two-dimensional (2D) transesophageal echocardiography (TEE; device size=1.1177×TEE defect size+3.5297; R=0.84; p<0.010).
Conclusion: EA and ER did not show a significant correlation with sex and defect size in our study. Patients older than 40 had a significantly lower ER compared to younger patients. Device size was strongly correlated with defect size measured with TEE.