Objectives: To assess accuracy and repeatability of the CASIA swept-source optical coherence tomographer (SS-OCT) in measuring contact lens (CL) radii of curvature and thicknesses compared with verified CL parameters and to investigate intrasession and intersession repeatability of the CASIA SS-OCT in measuring epithelial and total corneal thicknesses.

Methods: Rigid CLs with varying radii of curvature (front, rf; back, rb) and thicknesses were imaged with the CASIA SS-OCT across two sessions. Contact lens parameters were measured from processed images using an automated MATLAB program and were compared with parameters verified using standard techniques. Central epithelial and total corneal thicknesses of 43 normal subjects were measured to assess intrasession and intersession repeatability of the CASIA SS-OCT.

Results: No significant differences (P>0.05) were found in rigid CL rf and central and peripheral thickness measurements compared with verified CL parameters. However, the rb values obtained from the CASIA SS-OCT were consistently 0.2 mm flatter than the verified rb values (P<0.001) along horizontal and vertical meridians. Bland-Altman analysis revealed excellent intrasession agreement with mean differences of 0.004 and 0.003 mm for rf and rb, 0.000 mm for CL thickness and 0.372 and 0.395 μm for epithelial and total corneal thicknesses, respectively. Similarly, mean intersession differences of 0.020 and 0.000 mm for rf and CL thickness and 0.100 and 0.984 μm for epithelial and total corneal thicknesses were found, respectively. Ninety-five percentage confidence intervals across one or two sessions indicated insignificant undermeasurement or overmeasurement for CL rf, and corneal thicknesses, but significant bias toward overmeasurement for CL rb was found across two sessions.

Conclusions: The CASIA SS-OCT produces accurate measurements of CL parameters compared with verified values. Inbuilt distortion correction in this instrument necessitated no further correction of scanned images and provided high intrasession and intersession repeatability in measuring both CLs and corneal thicknesses. Further investigation of discrepancies in rb measurements is warranted.

Download full-text PDF

Source
http://dx.doi.org/10.1097/ICL.0000000000000417DOI Listing

Publication Analysis

Top Keywords

repeatability casia
12
casia ss-oct
12
accuracy repeatability
8
swept-source optical
8
optical coherence
8
coherence tomographer
8
ss-oct measuring
8
contact lens
8
radii curvature
8
compared verified
8

Similar Publications

Purpose: To investigate the accuracy of three distinct postoperative toric intraocular lens (TIOL) rotational stability measurement methods.

Setting: Department of Ophthalmology and Optometry, Medical University of Vienna, Austria.

Design: Single-center, prospective, interventional clinical trial.

View Article and Find Full Text PDF

Purpose: To evaluate and compare the repeatability of corneal densitometry (CD) measurements obtained using both an anterior-segment optical coherence tomography (AS-OCT) device and a Scheimpflug camera system, while also assessing the level of agreement. The study also sought to investigate the correlation of CD with age, gender, and central corneal thickness (CCT) in normal eyes.

Methods: CD measurements were obtained using the Casia 2 and the Pentacam AXL Wave.

View Article and Find Full Text PDF

Purpose: This study aims to evaluate the repeatability of anterior segment optical coherence tomography (AS-OCT) in diverse ocular surface disorder (OSD) cohorts, exploring various anterior segment parameters and their accuracy in different disease groups.

Methods: A total of 239 participants across six distinct OSD groups and healthy controls underwent nonmydriatic AS-OCT imaging using the Tomey CASIA 2 device. Anterior segment parameters including anterior chamber depth, width, angle metrics, corneal thickness, keratometry, lens vault, and others were meticulously assessed.

View Article and Find Full Text PDF

Background: To evaluate repeatability and agreement in measurements of total corneal astigmatism (TCA) in keratoconic eyes, using four optical coherence tomography (OCT)-based devices: Anterion, Casia SS-1000, IOLMaster 700, and MS-39.

Methods: Three consecutive measurements were taken with each device in 136 eyes. TCA values were converted into components J and J.

View Article and Find Full Text PDF

Introduction: Dry eye can compromise corneal astigmatism measurement repeatability during preoperative cataract surgery examination. No previous studies have analyzed the effectiveness of long-acting 3% diquafosol sodium (LA-DQS) on astigmatism measurement repeatability. This research assessed the effect of LA-DQS on astigmatism measurement repeatability in preoperative patients with cataract and short tear break-up time (TBUT) type dry eyes in both eyes of the same patient.

View Article and Find Full Text PDF

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!