A three-dimensional finite element model of the porcine lumbar spine (L1-L6) was used to assess the effect of spinal rod stiffness on lumbar biomechanics. The model was validated through a comparison with in vitro measurements performed on six porcine spine specimens. The validation metrics employed included intervertebral rotations and the nucleus pressure in the first instrumented intervertebral disc. The numerical results obtained suggest that rod stiffness values as low as 0.1 GPa are required to reduce the mobility gradient between the adjacent and instrumented segments and the nucleus pressures across the porcine lumbar spine significantly. Stiffness variations above this threshold value have no significant effect on spine biomechanics. For such low-stiffness rods, intervertebral rotations in the instrumented zone must be monitored closely in order to guarantee solid fusion. Looking ahead, the proposed model will serve to examine the transverse process hooks and variable stiffness rods in order to further smooth the transition between the adjacent and instrumented segments, while preserving the stability of the instrumented zone, which is needed for fusion.

Download full-text PDF

Source
http://dx.doi.org/10.1177/0954411917732596DOI Listing

Publication Analysis

Top Keywords

rod stiffness
12
porcine lumbar
12
spinal rod
8
lumbar biomechanics
8
finite element
8
element model
8
lumbar spine
8
intervertebral rotations
8
adjacent instrumented
8
instrumented segments
8

Similar Publications

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!