Introduction: Thalassemia could interfere with some assays for haemoglobin A (HbA) measurement, therefore, it is useful to be able to screen for thalassemia while measuring HbA. We used Capillarys 2 Flex Piercing (Capillarys 2FP) HbA programme to simultaneously measure HbA and screen for thalassemia.
Materials And Methods: Samples from 498 normal controls and 175 thalassemia patients were analysed by Capillarys 2FP HbA programme (Sebia, France). For method comparison, HbA was quantified by Premier Hb9210 (Trinity Biotech, Ireland) in 98 thalassaemia patients samples. For verification, HbA from eight thalassaemia patients was confirmed by IFCC reference method.
Results: Among 98 thalassaemia samples, Capillarys 2FP did not provide an HbA result in three samples with HbH due to the overlapping of HbBart's with HbA fraction; for the remaining 95 thalassaemia samples, Bland-Altman plot showed 0.00 ± 0.35% absolute bias between two systems, and a significant positive bias above 7% was observed only in two HbH samples. The HbA values obtained by Capillarys 2FP were consistent with the IFCC targets (relative bias below ± 6%) in all of the eight samples tested by both methods. For screening samples with alpha (α-) thalassaemia silent/trait or beta (β-) thalassemia trait, the optimal HbA cut-off values were ≤ 2.2% and > 2.8%, respectively.
Conclusions: Our results demonstrated the Capillarys 2FP HbA system could report an accurate HbA value in thalassemia silent/trait, and HbA value (≤ 2.2% for α-thalassaemia silent/trait and > 2.8% for β-thalassemia trait) and abnormal bands (HbH and/or HbBart's for HbH disease, HbF for β-thalassemia) may provide valuable information for screening.
Download full-text PDF |
Source |
---|---|
http://www.ncbi.nlm.nih.gov/pmc/articles/PMC5575649 | PMC |
http://dx.doi.org/10.11613/BM.2017.030704 | DOI Listing |
Scand J Clin Lab Invest
September 2018
a Department of Experimental Medicine and Surgery , "Tor Vergata" University Hospital, Rome , Italy.
Glycated hemoglobin (HbA) measurement provides the most important medium to long-term marker of time-averaged glycemic status. Its relationship to clinical outcome in diabetes has been convincingly demonstrated for both type 1 and type 2 diabetes. The main HbA measurement methods for clinical routine are as follows: ion-exchange chromatography; affinity chromatography, capillary electrophoresis, immunoassay and enzymatic methods.
View Article and Find Full Text PDFBiochem Med (Zagreb)
October 2017
Department of Laboratory Science, Second Affiliated Hospital of Guangzhou University of Chinese Medicine, Guangzhou, China.
Introduction: Thalassemia could interfere with some assays for haemoglobin A (HbA) measurement, therefore, it is useful to be able to screen for thalassemia while measuring HbA. We used Capillarys 2 Flex Piercing (Capillarys 2FP) HbA programme to simultaneously measure HbA and screen for thalassemia.
Materials And Methods: Samples from 498 normal controls and 175 thalassemia patients were analysed by Capillarys 2FP HbA programme (Sebia, France).
Biochem Med (Zagreb)
October 2016
Department of Laboratory Science, Second Affiliated Hospital of Guangzhou University of Chinese Medicine, Guangdong, China.
Introduction: Haemoglobin A (HbA) is widely used in the management of diabetes. Therefore, the reliability and comparability among different analytical methods for its detection have become very important.
Materials And Methods: A comparative evaluation of the analytical performances (precision, linearity, accuracy, method comparison, and interferences including bilirubin, triglyceride, cholesterol, labile HbA (LA), vitamin C, aspirin, fetal haemoglobin (HbF), and haemoglobin E (Hb E)) were performed on Capillarys 2 Flex Piercing (Capillarys 2FP) (Sebia, France), Tosoh HLC-723 G8 (Tosoh G8) (Tosoh, Japan), Premier Hb9210 (Trinity Biotech, Ireland) and Roche Cobas c501 (Roche c501) (Roche Diagnostics, Germany).
Am J Clin Pathol
June 2014
From the Department of Clinical Chemistry and Radio-immunology, Universitair Ziekenhuis Brussel, Brussels, Belgium.
Objectives: We evaluated the Sebia CAPILLARYS 2 Flex Piercing (Cap 2FP; Sebia, Lisses, France) for measurement of hemoglobin A1c (HbA1c) on venous and capillary blood samples.
Methods: We analyzed whole-blood samples and control materials with the Cap 2FP and Tosoh G8 (Tosoh Corporation, Tokyo, Japan). Capillary blood samples were analyzed on the Cap 2FP on different storage conditions and were compared with venous samples.
J Clin Lab Anal
January 2015
Department of Experimental Medicine and Surgery, University of Rome "Tor Vergata" Hospital, Rome, Italy.
Background: Glycated hemoglobin (HbA1c) provides a useful estimate of mean glycemia in patients with diabetes and is directly related to risks for diabetes complications. The aim of this study is to compare a capillary electrophoresis method and two high-performance liquid chromatography (HPLC) cation-exchange analyzers (Variant II (Bio-Rad Laboratories, Inc., Hercules, CA) and G8 (Tosoh Biosciences, San Francisco, CA)) to identify the most reliable method in Hb variants' presence.
View Article and Find Full Text PDFEnter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!