Severity: Warning
Message: file_get_contents(https://...@pubfacts.com&api_key=b8daa3ad693db53b1410957c26c9a51b4908&a=1): Failed to open stream: HTTP request failed! HTTP/1.1 429 Too Many Requests
Filename: helpers/my_audit_helper.php
Line Number: 176
Backtrace:
File: /var/www/html/application/helpers/my_audit_helper.php
Line: 176
Function: file_get_contents
File: /var/www/html/application/helpers/my_audit_helper.php
Line: 250
Function: simplexml_load_file_from_url
File: /var/www/html/application/helpers/my_audit_helper.php
Line: 1034
Function: getPubMedXML
File: /var/www/html/application/helpers/my_audit_helper.php
Line: 3152
Function: GetPubMedArticleOutput_2016
File: /var/www/html/application/controllers/Detail.php
Line: 575
Function: pubMedSearch_Global
File: /var/www/html/application/controllers/Detail.php
Line: 489
Function: pubMedGetRelatedKeyword
File: /var/www/html/index.php
Line: 316
Function: require_once
Objective: To quantify the long-term (>2y) effects of lower extremity (LE) neuroprostheses (NPs) for standing, transfers, stepping, and seated stability after spinal cord injury.
Design: Single-subject design case series with participants acting as their own concurrent controls, including retrospective data review.
Setting: Hospital-based clinical biomechanics laboratory with experienced (>20y in the field) research biomedical engineers, a physical therapist, and medical monitoring review.
Participants: Long-term (6.2±2.7y) at-home users (N=22; 19 men, 3 women) of implanted NPs for trunk and LE function with chronic (14.4±7.1y) spinal cord injury resulting in full or partial paralysis.
Interventions: Technical and clinical performance measurements, along with user satisfaction surveys.
Main Outcome Measures: Knee extension moment, maximum standing time, body weight supported by lower extremities, 3 functional standing tasks, 2 satisfaction surveys, NP usage, and stability of implanted components.
Results: Stimulated knee extension strength and functional capabilities were maintained, with 94% of implant recipients reporting being very or moderately satisfied with their system. More than half (60%) of the participants were still using their implanted NPs for exercise and function for >10min/d on nearly half or more of the days monitored; however, maximum standing times and percentage body weight through LEs decreased slightly over the follow-up interval. Stimulus thresholds were uniformly stable. Six-year survival rates for the first-generation implanted pulse generator (IPG) and epimysial electrodes were close to 90%, whereas those for the second-generation IPG along with the intramuscular and nerve cuff electrodes were >98%.
Conclusions: Objective and subjective measures of the technical and clinical performances of implanted LE NPs generally remained consistent for 22 participants after an average of 6 years of unsupervised use at home. These findings suggest that implanted LE NPs can provide lasting benefits that recipients value.