Background: Thoracic endovascular aortic repair (TEVAR) has dramatically expanded treatment options for patients with thoracic aortic pathology. The interaction between endografts and the dynamic anatomy of the thoracic aorta is not well characterized for repetitive physiologic stressors and subsequent issues related to long-term durability. Through three-dimensional (3D) modeling we sought to quantify cardiac-induced aortic deformation before and after TEVAR to assess the impact of endografts on dynamic aortic anatomy.
Methods: Eight patients with acute (n = 4) or chronic (n = 3) type B dissections, or chronic arch aneurysm (n = 1), underwent TEVAR with a single (n = 5) or multiple (n = 3) Gore C-TAG(s). Cardiac-resolved thoracic CT images were acquired pre- and post-TEVAR. 3D models of thoracic aorta and branch vessels were constructed in systole and diastole. Axial length, mean, and peak curvature of the ascending aorta, arch, and stented lumens were computed from the aortic lumen centerline, delineated with branch vessel landmarks. Cardiac-induced deformation was computed from mid-diastole to end-systole.
Results: Pre-TEVAR, there were no significant cardiac-induced changes for aortic axial length or mean curvature. Post-TEVAR, the ascending aorta increased in axial length (2.7 ± 3.1%, P < 0.05) and decreased in mean curvature (0.38 ± 0.05 → 0.36 ± 0.05 cm, P < 0.05) from diastole to systole. From pre- to post-TEVAR, axial length change increased in the ascending aorta (P < 0.02), mean curvature decreased in the arch and stented aorta (P < 0.03), and peak curvature decreased in the stented aorta (P < 0.05).
Conclusions: TEVAR for a range of indications not only causes direct geometric changes to the stented aorta but also results in dynamic changes to the ascending and stented aorta. In our cohort, endograft placement straightens the stented aorta and mutes cardiac-induced bending due to longitudinal stiffness. This is compensated by greater length and curvature changes from diastole to systole in the ascending aorta, relative to pre-TEVAR.
Download full-text PDF |
Source |
---|---|
http://dx.doi.org/10.1016/j.avsg.2017.07.033 | DOI Listing |
Sci Rep
January 2025
Department of Engineering Mechanics, KTH Royal Institute of Technology, Stockholm, Sweden.
Aneurysm rupture is a life-threatening event, yet its underlying mechanisms remain largely unclear. This study investigated the fracture properties of the thoracic aneurysmatic aorta (TAA) using the symmetry-constraint Compact Tension (symconCT) test and compared results to native and enzymatic-treated porcine aortas' tests. With age, the aortic stiffness increased, and tissues ruptured at lower fracture energy [Formula: see text].
View Article and Find Full Text PDFJ Med Genet
January 2025
Center of Medical Genetics, University of Antwerp and Antwerp University Hospital, Antwerp, Belgium.
Background: Individuals harbouring pathogenic variants are at risk for aneurysms/dissections throughout the arterial tree. Based on prior reports of sex differences in thoracic aortic aneurysm/dissection, we investigated the sexual dimorphism for vascular events in variant-harbouring patients.
Methods: We analysed two large pedigrees comprising 84 individuals segregating pathogenic missense variants affecting the same p.
A A Pract
January 2025
Department of Anesthesiology, University of Kansas Medical Center, Kansas City, Kansas.
Transcatheter aortic valve replacement (TAVR) is a common treatment for severe aortic stenosis (AS), but it carries the risk of severe complications, including device embolization. We present a case of a TAVR valve embolization into the left ventricular outflow tract (LVOT), diagnosed with transesophageal echocardiography (TEE) shortly after device deployment. The dislodged valve was successfully retrieved from the LVOT into the aorta, flattened, and stabilized with a thoracic endovascular aneurysm repair (TEVAR) stent, enabling the successful implantation of a new TAVR valve.
View Article and Find Full Text PDFArterioscler Thromb Vasc Biol
January 2025
Division of Cardiology, Department of Medicine, University of Washington (S.S., S.J., N.S., C.Y.L., L.L., D.A.D.).
Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!