Background And Purpose: The Woven EndoBridge device was recently introduced for the intrasaccular treatment of wide-neck aneurysms without the need for adjunctive devices. We present our results of the primary treatment of ruptured aneurysms with the Woven EndoBridge regardless of location or neck size.
Materials And Methods: Between February 2015 and April 2017, 100 ruptured aneurysms were selectively treated with the Woven EndoBridge. No supporting stents or balloons were used. There were 71 women treated (mean patient age, 59 years; median age, 60 years; range, 23-82 years).
Results: The mean aneurysm size was 5.6 mm (range, 3-13 mm), and 42 aneurysms were ≤4 mm. Sixty-six aneurysms (66%) had a wide neck, defined as ≥4 mm or a dome-neck ratio ≤1.5. There was 1 procedural rupture without sequelae. In 9 patients (9%), thromboembolic complications occurred. One poor grade patient died; neurologic deficits remained in 3. Overall treatment-related morbidity-mortality was 4% (4 of 100; 95% CI, 1.2%-10.2%).Two of 100 aneurysms were initially incompletely occluded and were additionally treated early after initial intervention. Of 80 eligible patients, 74 (93%) had 3-month angiographic follow-up. Fifty-four aneurysms (73%) were completely occluded, 17 (23%) had a small neck remnant, and 3 (4%) were incompletely occluded. One patient was additionally treated with a second Woven EndoBridge, and in 2 patients, additional treatment is scheduled. The overall reopening/retreatment rate was 6.8% (5 of 74; 95% CI, 2.6%-15.2%). There were no rebleeds during follow-up.
Conclusions: Treatment of small ruptured aneurysms with the Woven EndoBridge was safe and effective. The Woven EndoBridge proved to be a valuable alternative to coils without the need for stents or balloons.
Download full-text PDF |
Source |
---|---|
http://www.ncbi.nlm.nih.gov/pmc/articles/PMC7963751 | PMC |
http://dx.doi.org/10.3174/ajnr.A5371 | DOI Listing |
Neuroradiol J
January 2025
Department of Neurology, Neurosurgery & Radiology, University of Iowa Hospitals and Clinics, USA.
Background: The Woven EndoBridge 17 (WEB-17) is the latest advancement in the WEB device family. Comprehensive data on its occlusion rates, procedural complications, and mortality is lacking. This meta-analysis aimed to evaluate the efficacy and safety of the WEB-17 device in intracranial aneurysms (IAs).
View Article and Find Full Text PDFJ Neurointerv Surg
January 2025
Department of Neurosurgery, University at Buffalo Jacobs School of Medicine and Biomedical Sciences, Buffalo, New York, USA
Background: Early literature on the Woven EndoBridge (WEB) device reported 80-90% adequate aneurysm occlusion but low complete occlusion (40-55%). It is uncertain whether residual or recurrent aneurysms require re-treatment to prevent future rupture.
Objective: To systematically review the literature to meta-analyze occlusion and complication rates after re-treatment of these aneurysms.
J Cerebrovasc Endovasc Neurosurg
January 2025
Department of Neurosurgery, Tehran University of Medical Sciences, Tehran, Iran.
Objective: Treating wide-necked bifurcation aneurysms (WNBA) is challenging. Nevertheless, recent progress in endovascular techniques is promising. Woven EndoBridge devices (WEB) have exhibited outcomes comparable to conventional treatments like stent-assisted coiling (SAC) in treating aneurysms.
View Article and Find Full Text PDFInterv Neuroradiol
January 2025
Department of Interventional Neuroradiology, Austin Health, Heidelberg, Melbourne, Australia.
Background: Intrasaccular flow diversion using the woven endobridge device (WEB; MicroVention, Aliso Viejo, CA, USA) for the treatment of intracranial aneurysms has demonstrated large scale safety and efficacy. However, limitations arise from its structural configuration, restricting its application to specific aneurysm sizes and shapes.
Technique Overview: We introduce the CUPCAKE technique, a combination of conventional coiling followed by WEB intrasaccular flow disruption in select cases of atypical aneurysms with technically challenging morphology not typically treatable by WEB alone.
Transl Stroke Res
December 2024
Department of Neurological Surgery, Thomas Jefferson University Hospital, 901 Walnut street 3rd Floor, Philadelphia, PA, 19107, USA.
The Woven EndoBridge (WEB) device has become a prominent treatment for wide-neck bifurcation intracranial aneurysms since its FDA approval in 2018. However, the impact of anticoagulant therapy on its efficacy and patient outcomes remains underexplored. This study aims to evaluate the effects of postoperative anticoagulant use on aneurysm occlusion, retreatment rates, and functional outcomes following WEB device implantation.