Comparison of conventional reconstruction plate versus direct metal laser sintering plate: an in vitro mechanical characteristics study.

J Orthop Surg Res

Guangdong Provincial Key laboratory of Medical Biomechanics, Department of Anatomy, School of Basic Medicine Science, Southern Medical University, 1023 ShaTai Rd, Guangzhou, 510515, People's Republic of China.

Published: September 2017

Background: Additive manufacturing (AM) technology has helped to achieve several advances in the medical field, particularly as far as fabrication of implants is concerned. But the application of direct metal laser sintering (DMLS) bone plate is quite limited due to the indeterminate mechanical property. The purposes of this study were to characterize the biomechanical properties of the polished DMLS reconstruction plate and to compare these with the properties of commonly applied implants and to find whether the mechanical performance of DMLS plate meets the requirements for clinical application.

Methods: In this study, we fabricated two groups of plates by DMLS and computer numerical control (CNC) techniques. After that, we polished all samples and investigated their roughness, components, hardness, static bending, and torsional performance. Moreover, cyclic bending tests and fractographic analysis were conducted. Statistical comparisons of the group by means of monotonic test data were made, and a qualitative comparison was performed to assess failures in fatigue.

Results: We found no differences in surface roughness or components after polishing, but the DMLS plate hardness is 7.42% (p < 0.01) greater than that of the CNC plates. Compared with the CNC plates, the DMLS plate static bending and torsional performance were significantly greater. In a dynamic test, the DMLS plates survived 10, 10, 32,731, and 33,264 cycles under 0.6, 0.8, 0.9, and 1 kN cyclic loads, respectively, while the CNC plates survived 10, 10, 10, and 283,714 cycles.

Conclusions: These results indicate that the mechanical performances of the DMLS plate are stronger, and the strength under fatigue tests is sufficient. DMLS implant has great potential and may become a better choice for clinical use in the future. However, direct application of these AM instruments in the operating room requires further validation including animal and clinical experiment.

Download full-text PDF

Source
http://www.ncbi.nlm.nih.gov/pmc/articles/PMC5581440PMC
http://dx.doi.org/10.1186/s13018-017-0628-6DOI Listing

Publication Analysis

Top Keywords

reconstruction plate
8
direct metal
8
metal laser
8
laser sintering
8
dmls plate
8
roughness components
8
plate
6
dmls
5
comparison conventional
4
conventional reconstruction
4

Similar Publications

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!