A quick HPLC-ESI-MS/MS method was established for simultaneous determination of four major diterpenoids in Rabdosia japonica var.glaucocalyx, including glaucocalyxin A, oridonin, hebeirubesensin and enmenol. Analysis was performed on an Agilent ZORBAX SB-C18(4.6 mm×250 mm, 5 μm ) column eluted in a gradient program with methanol and water. The flow rate was 0.8 mL•min⁻¹. Multiple reaction monitoring (MRM) scanning mode was performed in negative ion switching mode to apply for the quantitative determination. The calibration curves for the above four compounds were linear in corresponding injection amount. The average recoveries of the compounds ranged from 92.40% to 105.9%, with RSDs of 1.7%-6.5%. The method is simple, rapid, accurate with good repeatability, which can provide a reference for overcalling evaluation the quality of R. japonica var.glaucocalyx. The result of cluster analysis- showed that the quality of R. japonica glaucocalyx var. greatly varied between areas and parts.
Download full-text PDF |
Source |
---|---|
http://dx.doi.org/10.4268/cjcmm20160215 | DOI Listing |