Optimizing a micro-computed tomography-based surrogate measurement of bone-implant contact.

J Orthop Res

Department of Cell & Molecular Medicine, Rush University Medical Center, 600 S. Paulina St., Suite 507, Chicago, Illinois, 60612.

Published: March 2018

Histology and backscatter scanning electron microscopy (bSEM) are the current gold standard methods for quantifying bone-implant contact (BIC), but are inherently destructive. Microcomputed tomography (μCT) is a non-destructive alternative, but attempts to validate μCT-based assessment of BIC in animal models have produced conflicting results. We previously showed in a rat model using a 1.5 mm diameter titanium implant that the extent of the metal-induced artefact precluded accurate measurement of bone sufficiently close to the interface to assess BIC. Recently introduced commercial laboratory μCT scanners have smaller voxels and improved imaging capabilities, possibly overcoming this limitation. The goals of the present study were to establish an approach for optimizing μCT imaging parameters and to validate μCT-based assessment of BIC. In an empirical parametric study using a 1.5 mm diameter titanium implant, we determined 90 kVp, 88 µA, 1.5 μm isotropic voxel size, 1600 projections/180°, and 750 ms integration time to be optimal. Using specimens from an in vivo rat experiment, we found significant correlations between bSEM and μCT for BIC with the manufacturer's automated analysis routine (r = 0.716, p = 0.003) or a line-intercept method (r = 0.797, p = 0.010). Thus, this newer generation scanner's improved imaging capability reduced the extent of the metal-induced artefact zone enough to permit assessment of BIC. © 2017 Orthopaedic Research Society. Published by Wiley Periodicals, Inc. J Orthop Res 36:979-986, 2018.

Download full-text PDF

Source
http://www.ncbi.nlm.nih.gov/pmc/articles/PMC5832531PMC
http://dx.doi.org/10.1002/jor.23716DOI Listing

Publication Analysis

Top Keywords

assessment bic
12
bone-implant contact
8
validate μct-based
8
μct-based assessment
8
15 mm diameter
8
diameter titanium
8
titanium implant
8
extent metal-induced
8
metal-induced artefact
8
improved imaging
8

Similar Publications

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!