Objective: To determine whether a new catheter design with a low-profile, open-ended rounded rather than pointed Foley catheter tip can reduce mucosal damage to the bladder of ewes.

Methods And Materials: Six ewes were randomly assigned to 1 of 2 indwelling urinary catheters-a 16 Fr Foley catheter or a 16 Fr open-tip CystoSure catheter. After 96 hours, all the animals were sacrificed and their bladder and urethra harvested for analysis.

Results: Image analysis of the bladder surfaces demonstrated a significant decrease in the percentage of bladder area covered by ulceration and inflammation in sheep with CystoSure catheters compared with Foley catheters (P < .002) as well as a trend toward less edema (P = .17). Macro-morphologic evaluations were confirmed with immunohistochemical markers of cell proliferation and inflammation.

Conclusion: In this pilot study, we were able to demonstrate that a new catheter design with an open-ended rounded rather than pointed tip and a reduced balloon base-to-tip profile may reduce mucosal damage to the bladder of ewes. Based on the findings from this trial, we believe this new catheter design with its low-profile, rounded tip may reduce bladder mucosal injury, which is a risk factor for catheter-associated urinary tract infections.

Download full-text PDF

Source
http://dx.doi.org/10.1016/j.urology.2017.08.020DOI Listing

Publication Analysis

Top Keywords

mucosal damage
8
indwelling urinary
8
foley catheter
8
bladder
5
preliminary evaluation
4
evaluation ovine
4
ovine bladder
4
bladder mucosal
4
damage associated
4
associated indwelling
4

Similar Publications

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!