  • The Vitamin D External Quality Assessment Scheme (DEQAS) now includes the National Institute of Standards and Technology (NIST), which uses a precise measurement procedure to assign accuracy-based target values for serum total 25-hydroxyvitamin D levels.
  • NIST analyzed 90 samples from various laboratories to compare their results against a consensus mean and assess biases for different testing methods like LC-MS/MS and HPLC.
  • The average difference between NIST values and the consensus mean was 5.6%, with some samples showing even greater discrepancies, highlighting the importance of accurate measurement standards in vitamin D testing.

Article Abstract

Until recently, the Vitamin D External Quality Assessment Scheme (DEQAS) assessed the performance of various assays for the determination of serum total 25-hydroxyvitamin D [25(OH)D] by using a consensus mean based on the all-laboratory trimmed mean (ALTM) of the approximately 1000 participants' results. Since October 2012, the National Institute of Standards and Technology (NIST), as part of the Vitamin D Standardization Program, has participated in DEQAS by analyzing the quarterly serum sample sets using an isotope dilution LC-tandem MS (ID LC-MS/MS) reference measurement procedure to assign an accuracy-based target value for serum total 25(OH)D. NIST has analyzed 90 DEQAS samples (18 exercises × 5 samples/exercise) to assign target values. The NIST-assigned values are compared with the ALTM and the biases assessed for various assays used by the participants, e.g., LC-MS/MS, HPLC, and several ligand-binding assays. The NIST-value assignment process and the results of the analyses of the 90 DEQAS samples are summarized. The absolute mean bias between the NIST-assigned values and the ALTM was 5.6%, with 10% of the samples having biases >10%. Benefits of the accuracy-based target values are presented, including for sample sets with high concentrations of 25(OH)D2 and 3-epi-25(OH)D3.

Download full-text PDF

Source
http://dx.doi.org/10.5740/jaoacint.17-0306DOI Listing

Publication Analysis

Top Keywords

vitamin external
8
external quality
8
quality assessment
8
assessment scheme
8
scheme deqas
8
serum total
8
sample sets
8
accuracy-based target
8
deqas samples
8
target values
8