Background: -butyl 2-cyanoacrylate (NBCA) remains the standard embolic agent for spinal dural arteriovenous fistula (SDAVF) treatment. Treatment of SDAVF with ethylene-vinyl alcohol copolymer (Onyx, ev3-Covidien, Irvine CA, USA) is currently not well established. Although several cases have reported the use of Onyx to embolize an intracranial dural arteriovenous fistula using a dual-lumen microballoon catheter, Onyx embolization of an SDAVF using a dual-lumen microballoon catheter has not been reported.

Case Description: We treated a 57-year-old man with an SDAVF using a dual-lumen microballoon catheter and buddy wire technique to perform transarterial Onyx embolization via the left sixth intercostal artery.

Conclusions: Onyx embolization using a dual-lumen microballoon catheter was effective. Furthermore, the buddy wire technique was useful for providing rigid support of the microcatheter in a narrow and tortuous intercostal artery.

Download full-text PDF

Source
http://www.ncbi.nlm.nih.gov/pmc/articles/PMC5551288PMC
http://dx.doi.org/10.4103/sni.sni_193_17DOI Listing

Publication Analysis

Top Keywords

dual-lumen microballoon
20
microballoon catheter
20
dural arteriovenous
12
arteriovenous fistula
12
buddy wire
12
wire technique
12
onyx embolization
12
spinal dural
8
ethylene-vinyl alcohol
8
alcohol copolymer
8

Similar Publications

Utilisation of the Scepter Mini dual-lumen balloon - An illustrative series.

Interv Neuroradiol

November 2023

Division of Neuroradiology, Joint Department of Medical Imaging, Toronto Western Hospital, University Health Network, Toronto, Ontario, Canada.

Background: Dual-lumen balloon microcatheters can aid in the safety and efficacy of endovascular embolisation of cerebrospinal vascular malformations. The Scepter Mini dual-lumen balloon is a novel device with a smaller profile than previous balloon microcatheters, opening up new indications not only in the treatment of cerebrospinal malformations but in various other neurovascular therapeutic and diagnostic scenarios.

Methods: Following institutional ethics review board approval, a retrospective review of our prospectively maintained database of cases employing the Scepter Mini dual-lumen microballoon catheter was conducted.

View Article and Find Full Text PDF

Initial experience with Scepter Mini dual lumen balloon for embolization of cerebrovascular diseases.

J Cerebrovasc Endovasc Neurosurg

June 2023

Department of Radiology, Diagnostic & Interventional Neuroradiology Section, King Fahad Medical City, Riyadh, Saudi Arabia.

Objective: Endovascular treatment of cerebrovascular diseases is often challenging due to small caliber, tortuous distal vessels. Several devices and techniques have evolved to overcome these challenges. Recently, a low profile dual lumen microballoon catheter, specifically designed for distal navigation is employed for neurovascular procedures.

View Article and Find Full Text PDF

Purpose: Dural arteriovenous fistulas (dAVF) account for approximately 10-15% of all intracranial arteriovenous abnormalities. dAVFs carry a significant risk of mortality, particularly in cases of acute hemorrhage, of up to 10%. A small proportion of these dAVFs are found in the anterior cranial fossa (ACF), of which the rate of hemorrhage can be as high as up to 91%.

View Article and Find Full Text PDF

Background: Balloon-assisted techniques can improve the endovascular treatment of cerebrospinal vascular malformations. The aim of this study was to report the first clinical multicenter experience with the new Scepter Mini dual-lumen microballoon catheter.

Methods: Patients with cerebral or spinal vascular malformations treated with the Scepter Mini at seven European neurovascular centers were retrospectively reviewed.

View Article and Find Full Text PDF

Background: The embolization technique can have significant impact on the success of endovascular embolization.

Objective: To evaluate the feasibility, embolization characteristics, and embolization extent with a newly developed extra-small micro-balloon catheter in an in vivo and in an in vitro embolization model in comparison with standard microcatheter embolizations.

Materials And Methods: Twenty-eight embolization procedures were performed in the porcine rete mirabile (RM) and in an in vitro embolization model, using either an extra-small (distal outer diameter 1.

View Article and Find Full Text PDF

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!