Validity of a smartphone protractor to measure sagittal parameters in adult spinal deformity.

Spine J

Department of Orthopedic Surgery, Yawkey Center for Outpatient Care, Massachusetts General Hospital, 55 Fruit St, Suite 3A, Boston, MA 02114, USA.

Published: October 2017

Background Context: Smartphones have become an integral tool in the daily life of health-care professionals (Franko 2011). Their ease of use and wide availability often make smartphones the first tool surgeons use to perform measurements. This technique has been validated for certain orthopedic pathologies (Shaw 2012; Quek 2014; Milanese 2014; Milani 2014), but never to assess sagittal parameters in adult spinal deformity (ASD). This study was designed to assess the validity, reproducibility, precision, and efficiency of using a smartphone protractor application to measure sagittal parameters commonly measured in ASD assessment and surgical planning.

Purpose: This study aimed to (1) determine the validity of smartphone protractor applications, (2) determine the intra- and interobserver reliability of smartphone protractor applications when used to measure sagittal parameters in ASD, (3) determine the efficiency of using a smartphone protractor application to measure sagittal parameters, and (4) elucidate whether a physician's level of experience impacts the reliability or validity of using a smartphone protractor application to measure sagittal parameters in ASD.

Study Design/setting: An experimental validation study was carried out.

Methods: Thirty standard 36″ standing lateral radiographs were examined. Three separate measurements were performed using a marker and protractor; then at a separate time point, three separate measurements were performed using a smartphone protractor application for all 30 radiographs. The first 10 radiographs were then re-measured two more times, for a total of three measurements from both the smartphone protractor and marker and protractor. The parameters included lumbar lordosis, pelvic incidence, and pelvic tilt. Three raters performed all measurements-a junior level orthopedic resident, a senior level orthopedic resident, and a fellowship-trained spinal deformity surgeon. All data, including the time to perform the measurements, were recorded, and statistical analysis was performed to determine intra- and interobserver reliability, as well as accuracy, efficiency, and precision. Statistical analysis using the intra- and interclass correlation coefficient was calculated using R (version 3.3.2, 2016) to determine the degree of intra- and interobserver reliability.

Results: High rates of intra- and interobserver reliability were observed between the junior resident, senior resident, and attending surgeon when using the smartphone protractor application as demonstrated by high inter- and intra-class correlation coefficients greater than 0.909 and 0.874 respectively. High rates of inter- and intraobserver reliability were also seen between the junior resident, senior resident, and attending surgeon when a marker and protractor were used as demonstrated by high inter- and intra-class correlation coefficients greater than 0.909 and 0.807 respectively. The lumbar lordosis, pelvic incidence, and pelvic tilt values were accurately measured by all three raters, with excellent inter- and intra-class correlation coefficient values. When the first 10 radiographs were re-measured at different time points, a high degree of precision was noted. Measurements performed using the smartphone application were consistently faster than using a marker and protractor-this difference reached statistical significance of p<.05.

Conclusions: Adult spinal deformity radiographic parameters can be measured accurately, precisely, reliably, and more efficiently using a smartphone protractor application than with a standard protractor and wax pencil. A high degree of intra- and interobserver reliability was seen between the residents and attending surgeon, indicating measurements made with a smartphone protractor are unaffected by an observer's level of experience. As a result, smartphone protractors may be used when planning ASD surgery.

Download full-text PDF

Source
http://dx.doi.org/10.1016/j.spinee.2017.06.014DOI Listing

Publication Analysis

Top Keywords

smartphone protractor
36
sagittal parameters
24
measure sagittal
20
protractor application
20
intra- interobserver
16
validity smartphone
12
protractor
12
spinal deformity
12
application measure
12
interobserver reliability
12

Similar Publications

Minimally Invasive Derotational Osteotomy of Long Bones: Smartphone Application Used to Improve the Accuracy of Correction.

J Clin Med

February 2023

Department of Orthopedic Surgery, School of Medicine, Kyungpook National University, Kyungpook National University Hospital, Daegu 41944, Republic of Korea.

Correction of rotational malalignments caused by fractures is essential as it may cause pain and gait disturbances. This study evaluated the intraoperative use of a smartphone application (SP app) to measure the extent of corrective rotation in patients treated using minimally invasive derotational osteotomy. Intraoperatively, two parallel 5 mm Schanz pins were placed above and below the fractured/injured site, and derotation was performed manually after percutaneous osteotomy.

View Article and Find Full Text PDF

Introduction: The degree of chordee associated with hypospadias impacts operative management. Unfortunately, poor inter-observer reliability in assessing chordee by multiple methods in vitro has been demonstrated. This variability may be related to the fact that chordee is not a discrete angle, but rather an arc-like curvature similar to that of a banana.

View Article and Find Full Text PDF

Purpose: To evaluate the accuracy of needle placement using a three-dimensional (3D) augmented reality (AR) protractor on smartphones (AR Puncture).

Materials And Methods: An AR protractor that can be rotated in three directions against the CT plane with angle guidance lines for smartphones was developed. The protractor center can be adjusted to an entry point by manually moving the smartphone with the protractor center fixed at the center of the screen (Fix-On-Screen) or by image tracking with a printed QR code placed at an entry point (QR-Tracking).

View Article and Find Full Text PDF

The protractor-based goniometer is the standard instrument used to measure finger range of motion (ROM). However, the method is often complicated to apply and places a burden on the investigator. Here, we developed a new method for finger ROM measurement using a smartphone.

View Article and Find Full Text PDF

sEVD-smartphone-navigated placement of external ventricular drains.

Acta Neurochir (Wien)

March 2020

Department of Neurosurgery, Inselspital, Bern University Hospital, University of Bern, Freiburgstrasse 10, 3010, Bern, Switzerland.

Background: Currently, the trajectory for insertion of an external ventricular drain (EVD) is mainly determined using anatomical landmarks. However, non-assisted implantations frequently require multiple attempts and are associated with EVD malpositioning and complications. The authors evaluated the feasibility and accuracy of a novel smartphone-guided, angle-adjusted technique for assisted implantations of an EVD (sEVD) in both a human artificial head model and a cadaveric head.

View Article and Find Full Text PDF

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!