Pullout strength of 2.0 mm cancellous and cortical screws in synthetic bone.

Vet Surg

Department of Mathematical Sciences, Clemson University, Clemson, South Carolina.

Published: November 2017

Objective: To determine whether 2.0 mm cancellous screws are superior to 2.0 mm cortical screws when inserted into cancellous and bicortical bone.

Study Design: Biomechanical study.

Methods: The 2.0 mm cancellous screws and 2.0 mm cortical screws were inserted according to the recommended guidelines in synthetic cancellous and bicortical blocks. Fifteen screw-block constructs per group were tested to failure in axial pullout. Axial pullout strength and yield strength were calculated. Data were analyzed using a one-way ANOVA.

Results: The 2.0 mm cortical screws achieved lower axial pullout strength than 2.0 mm cancellous screws in cancellous blocks. The 2.0 mm cortical screws achieved greater pullout strength than 2.0 mm cancellous screws in bicortical blocks.

Conclusion: The 2.0 mm cancellous screws may offer a biomechanical advantage in bone with thin cortices (<1 mm thick), whereas 2.0 mm cortical screws may be preferred in cortical bone with cortices measuring at least 1 mm in thickness.

Download full-text PDF

Source
http://dx.doi.org/10.1111/vsu.12692DOI Listing

Publication Analysis

Top Keywords

cortical screws
20
cancellous screws
20
pullout strength
16
strength cancellous
12
axial pullout
12
screws
10
cancellous
9
screws inserted
8
cancellous bicortical
8
screws achieved
8

Similar Publications

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!