Preventing Retained Central Venous Catheter Guidewires: A Randomized Controlled Simulation Study Using a Human Factors Approach.

Anesthesiology

From the Division of Anaesthesia, University of Cambridge, Addenbrooke's Hospital, Cambridge, United Kingdom (M.Z.A.M., D.K.M.); Department of Anesthesia and Perioperative Medicine, Medical University of South Carolina, Charleston, South Carolina (K.R.C.); Cambridge University Hospitals, Addenbrooke's Hospital, Cambridge, United Kingdom (A.K.G.); and Critical Care Department, The Queen Elizabeth Hospital, Kings Lynn, United Kingdom (P.J.Y.).

Published: October 2017

Background: Retained central venous catheter guidewires are never events. Currently, preventative techniques rely on clinicians remembering to remove the guidewire. However, solutions solely relying upon humans to prevent error inevitably fail. A novel locked procedure pack was designed to contain the equipment required for completing the procedure after the guidewire should have been removed: suture, suture holder, and antimicrobial dressings. The guidewire is used as a key to unlock the pack and to access the contents; thereby, the clinician must remove the guidewire from the patient to complete the procedure.

Methods: A randomized controlled forced-error simulation study replicated catheter insertion. We created a retained guidewire event and then determined whether clinicians would discover it, comparing standard practice against the locked pack.

Results: Guidewires were retrieved from 2/10 (20%) standard versus 10/10 (100%) locked pack, n = 20, P < 0.001. In the locked pack group, participants attempted to complete the procedure; however, when unable to access the contents, this prompted a search for the key (guidewire). Participants discovered the guidewire within the catheter lumen, recovered it, utilized it to unlock the pack, and finish the procedure. A structured questionnaire reported that the locked pack also improved subjective safety of central venous catheter insertion and allowed easy disposal of the sharps and guidewire (10/10).

Conclusions: The locked pack is an engineered solution designed to prevent retained guidewires. Utilizing forced-error simulation testing, we have determined that the locked pack is an effective preventative device and is acceptable to clinicians for improving patient safety.

Download full-text PDF

Source
http://dx.doi.org/10.1097/ALN.0000000000001797DOI Listing

Publication Analysis

Top Keywords

locked pack
20
central venous
12
venous catheter
12
retained central
8
catheter guidewires
8
randomized controlled
8
simulation study
8
guidewire
8
remove guidewire
8
pack
8

Similar Publications

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!