Background: Glucose sensors, such as FreeStyle® Libre, are innovative medical devices developed for diabetes patients as a replacement for classic glucose meters, ensuring continuous glucose monitoring without the disadvantage of regular skin finger pricks.

Objectives: To report several cases of allergic contact dermatitis caused by FreeStyle® Libre, and to report on isobornyl acrylate as a culprit allergen.

Patients And Methods: Fifteen patients presented with allergic contact dermatitis caused by FreeStyle® Libre. All but 1 were patch tested with a baseline series, and with pieces and/or ultrasonic bath extracts of (the adhesive part of) the glucose sensor. Isobornyl acrylate was patch tested, in various concentrations and vehicles, in 13 patients. Gas chromatography-mass spectrometry (GC-MS) of the sensors was performed.

Results: All patients reacted to the adhesive part of the sensor, and 12 patients were shown to be sensitized to isobornyl acrylate. Simultaneous reactions to other allergens were rarely observed. GC-MS showed the presence of isobornyl acrylate in the sensors.

Conclusions: Cases of allergic contact dermatitis caused by FreeStyle® Libre are increasingly being observed, and isobornyl acrylate is a relevant culprit allergen. Cross-reactivity to other acrylates was infrequently observed, but other, hitherto unidentified, contact allergens may still be present in the device.

Download full-text PDF

Source
http://dx.doi.org/10.1111/cod.12866DOI Listing

Publication Analysis

Top Keywords

isobornyl acrylate
24
freestyle® libre
20
allergic contact
16
contact dermatitis
16
dermatitis caused
16
caused freestyle®
12
glucose sensor
8
cases allergic
8
patch tested
8
isobornyl
6

Similar Publications

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!