Objectives: The purpose of this study was to assess the validity of the ActiGraph GT3X+ (GT3X+) and the BodyMedia SenseWear Armband (SWA) to estimate energy expenditure (EE) during physical activity and field sport movements.

Design: Criterion validity.

Methods: Twenty-six active adults completed a single 90min session involving alternating intervals of exercise (5min) and recovery (10min). Exercise involved walking (4km/h), jogging (8km/h), running (12km/h) or a sport-simulated circuit (three intervals). Participants wore two triaxial accelerometers (GT3X+ and SWA) and a portable gas analyser (MetaMax 3B), used as the criterion measure.

Results: Total EE was significantly underestimated (p<0.01) by the GT3X+ (mean bias±SD: -374.5±132.84kJ; % difference=-29.3%) and SWA (-244.3±148.0kJ; -18.2%). Overestimations were made by both accelerometers during the walk (GT3X+: 27.4±30.8kJ; SWA: 32.1±15.4kJ) and jog (38.0±30.0kJ; 34.5±31.6kJ). Underestimations were evident during the run (-41.2±25.1kJ; -43.8±33.5kJ) and circuit (C1: GTX+: -127.2±41.6kJ; SWA: -86.1±40.2kJ). Error of estimation increased in magnitude as the intensity of exercise increased (GT3X+: 40.8-143.0kJ; SWA: 35.5-102.0kJ).

Conclusions: The ActiGraph GT3X+ and BodyMedia SWA do not provide valid EE estimates across a range of exercise modalities and intensities when compared to a criterion measure. Poor accuracy and large precision errors, particularly during high intensity and intermittent movement patterns, suggest these devices have limitations and should be used cautiously in the field.

Download full-text PDF

Source
http://dx.doi.org/10.1016/j.jsams.2017.07.022DOI Listing

Publication Analysis

Top Keywords

validity actigraph
8
actigraph gt3x+
8
gt3x+ bodymedia
8
bodymedia sensewear
8
sensewear armband
8
estimate energy
8
energy expenditure
8
expenditure physical
8
physical activity
8
gt3x+
4

Similar Publications

We propose the addition of usability validation to the extended V3 framework, now "V3+", and describe a pragmatic approach to ensuring that sensor-based digital health technologies can be used optimally at scale by diverse users. Alongside the original V3 components (verification; analytical validation; clinical validation), usability validation will ensure user-centricity of digital measurement tools, paving the way for more inclusive, reliable, and trustworthy digital measures within clinical research and clinical care.

View Article and Find Full Text PDF

Proof-of-concept for integrating multimodal digital health assessments into lifestyle interventions for older adults with dementia risk factors.

J Behav Med

January 2025

Department of Psychiatry, Center for Health Outcomes and Interdisciplinary Research, Massachusetts General Hospital, One Bowdoin Square, 1st Floor, Suite 100, Boston, MA, 02114, USA.

Multimodal digital health assessments overcome the limitations of patient-reported outcomes by allowing for continuous and passive monitoring but remain underutilized in older adult lifestyle interventions for brain health. Therefore, we aim to (1) report ecological momentary assessment (EMA) and ActiGraph adherence among older adults during a lifestyle intervention; and (2) use dynamic data collected via EMA and ActiGraph to examine person-specific patterns of mindfulness, steps, and sleep throughout the intervention. We analyzed EMA and ActiGraph data from a pilot study of the 8-week My Healthy Brain program (N = 10) lifestyle group for older adults (60+) with subjective cognitive decline.

View Article and Find Full Text PDF

Comparing Five Generations of ActiGraph Devices using an Orbital Shaker.

Med Sci Sports Exerc

January 2025

Energy Metabolism Section, National Institute of Diabetes, Digestive and Kidney Diseases, Diabetes, Endocrinology, and Obesity Branch, National Institutes of Health (NIH), Bethesda, MD.

Introduction: ActiGraph accelerometers are used extensively to objectively assess physical activity, sedentary behavior, and sleep. Here, we present an objective validation of five generations of ActiGraph sensors to characterize potential differences in output arising from changes to hardware or firmware.

Methods: An orbital shaker generated accelerations from 0 to 3700 milli-g in a randomized order to test the wGT3X-BT, GT9X, CentrePoint Insight Watch (CPIW) 1.

View Article and Find Full Text PDF

Purpose: This paper describes the data collection and management methods for the Cancer Prevention Study-3 (CPS-3) Accelerometry Substudy, a nested cohort of device-based physical activity and sedentary time data.

Participants: US-based CPS-3 participants (initially enrolled 2006-2013) who completed the 2018 follow-up survey and had a valid email address were invited to the Accelerometry Substudy (n=109 780). Among the 23 111 participants who registered and were shipped an ActiGraph GT3X+ accelerometer, 21 219 participants returned the device with a complete wear log (91.

View Article and Find Full Text PDF

Goal And Aims: One challenge using wearable sensors is nonwear time. Without a nonwear (e.g.

View Article and Find Full Text PDF

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!