Purpose: To evaluate the intraoperator repeatability and interoperator reproducibility of two Scheimpflug-Placido anterior corneal analyzers (Sirius; Costruzione Strumenti Oftalmici, Florence, Italy, and TMS-5; Tomey, Nagoya, Japan) for the measurement of posterior corneal surface in normal eyes and the agreement between the two devices.

Methods: The current prospective study investigated 55 right eyes of 55 healthy patients in a random order. The posterior steep keratometry (Ks), flat keratometry (Kf), mean keratometry (Km), and astigmatism were randomly measured by two independent experienced operators using the Sirius and TMS-5 Scheimpflug-Placido systems. Three consecutive measurements were obtained. Vector analysis was used for astigmatism. To assess intraoperator repeatability and interoperator reproducibility, within-subject standard deviation (Sw) and test-retest repeatability (TRT, 2.77 Sw) were calculated. The Bland-Altman plots and 95% limits of agreement (LoA) were used to evaluate the agreement between the two systems.

Results: In the Sirius Scheimpflug-Placido system, the Sw and TRT of keratometry and astigmatism were less than 0.04 and 0.12 diopters (D), respectively. The interoperator Sw and TRT were less than 0.03 and 0.07 D, respectively. In the TMS-5 Scheimpflug-Placido system, the intraoperator Sw and TRT did not exceed 0.03 and 0.08 D, respectively. The interoperator Sw and TRT were no greater than 0.02 and 0.05 D, respectively. All parameters showed minor fluctuations in the 95% LoA.

Conclusions: Both Scheimpflug-Placido systems show high repeatability and reproducibility for posterior corneal surface measurement. A high concordance between the two devices suggests interchangeable use in normal eyes. [J Refract Surg. 2017;33(8):524-530.].

Download full-text PDF

Source
http://dx.doi.org/10.3928/1081597X-20170606-01DOI Listing

Publication Analysis

Top Keywords

posterior corneal
12
corneal surface
12
repeatability reproducibility
8
scheimpflug-placido anterior
8
anterior corneal
8
corneal analyzers
8
surface measurement
8
intraoperator repeatability
8
repeatability interoperator
8
interoperator reproducibility
8

Similar Publications

Purpose: The purpose of this study was to report the management of chemoimmunotherapy-resistant ocular surface squamous neoplasia (OSSN) with iodine-125 (I-125) brachytherapy.

Methods: A 36-year-old man presented to the clinic with biopsy-proven OSSN that covered ∼70% of the corneal surface and extended to the 6 o'clock position of the inferior limbus of the OS. The visual acuity was 20/20 in the OD and 20/40 in the affected OS.

View Article and Find Full Text PDF

Trans-iris vertical intrascleral haptic fixation of IOL: The bypass technique in eyes with megalocornea.

Indian J Ophthalmol

December 2024

Cornea, Cataract and Refractive Surgery Services, Dr Rajendra Prasad Centre for Ophthalmic Sciences, All India Institute of Medical Sciences, New Delhi, India.

Intrascleral haptic fixation of intraocular lens (IOL) is an extremely useful technique to provide visual rehabilitation in eyes with inadequate capsular support. It requires exteriorization of haptics along with tucking of haptics in the scleral groove preferably and conventionally in the horizontal meridian. In eyes with large corneal diameter, there is difficulty in tucking enough length of the haptics into the intrascleral groove, carrying the risk of slippage of haptics and decentration of IOL.

View Article and Find Full Text PDF

Objectives: To demonstrate corneal remodeling after corneal allograft intrastromal ring segment (CAIRS) with an anterior-segment optical coherence tomography (AS-OCT).

Design: A prospective observational single-center study.

Methods: This observational study included keratoconus patients who underwent CAIRS implantation into a stromal tunnel.

View Article and Find Full Text PDF

Predicting corneal decompensation in Fuchs endothelial corneal dystrophy with Scheimpflug tomography and clinical parameters.

Indian J Ophthalmol

January 2025

Department of Ophthalmology, Université Paris Cité, Cochin Hospital, Assistance Publique-Hôpitaux de Paris, Paris, France.

Purpose: This study aims to evaluate the efficacy of various tomographic indices, both established and novel, in predicting endothelial decompensation leading to either spontaneous corneal transplantation or transplantation following cataract surgery in patients with Fuchs endothelial corneal dystrophy (FECD).

Methods: In this cross-sectional, retrospective study, we reviewed the files of 93 eyes from 54 FECD patients undergoing regular follow-up. We recorded clinical metrics such as morning visual disturbance (MVD) and corrected distance visual acuity.

View Article and Find Full Text PDF

Introduction: The aim of our study was to assess the outcome of Gore-Tex sutures in minimally invasive scleral fixation of subluxated posterior chamber intraocular lenses (PCIOLs) and to demonstrate a method for validating the lens position.

Methods: Retrospective study of patients who underwent lasso in-the-bag scleral fixation of a subluxated PCIOL using the snare technique with Gore-Tex suture from 2019 to 2021 in a single tertiary medical center. Functional outcome was analyzed by clinical assessment, and anatomical outcome, by ultrasound biomicroscopy (UBM).

View Article and Find Full Text PDF

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!