Purpose: Pedicle subtraction osteotomy (PSO) is performed to treat rigid, sagittal spinal deformities, but high rates of implant failure are reported. Anterior lumbar interbody fusion has been proposed to reduce this risk, but biomechanical investigation is lacking. The goal of this study was to quantify the (1) destabilizing effects of a lumbar osteotomy and (2) contribution of anterior lumbar interbody fusion (ALIF) at the lumbosacral junction as recommended in literature.

Methods: Fourteen fresh human thoracolumbosacral spines (T12-S1) were tested in flexion-extension (FE), lateral bending (LB), and axial rotation (AR). Bilateral pedicle screws/rods (BPS) were inserted at T12-S1, cross connectors (CC) at T12-L1 and L5-S1, and anterior interbody spacers (S) at L4-5 and L5-S1. In one group, PSO was performed in seven specimens at L3. All specimens were sequentially tested in (1) Intact; (2) BPS; (3) BPS + CC; (4) BPS + S; and (5) BPS + S + CC; a second group of seven spines were tested in the same sequence without PSO. Mixed-model ANOVA with repeated measures was performed (p ≤ 0.05).

Results: At the osteotomy site (L2-L4), in FE, BPS, BPS + CC, BPS + S, BPS + CC + S reduced motion to 11.2, 12.9, 10.9, and 11.4%, respectively, with significance only found in BPS and BPS + S construction (p ≤ 0.05). All constructs significantly reduced motion across L2-L4 in the absence of PSO, across all loading modes (p ≤ 0.05). PSO significantly destabilized L2-L4 axial rotational stability, regardless of operative construction (p ≤ 0.05). Across L4-S1 and L2-S1, all instrumented constructs significantly reduced motion, in both PSO- and non-PSO groups, during all loading modes (p ≤ 0.05).

Conclusions: These findings suggest anterior interbody fusion minimally immobilizes motion segments, and interbody devices may primarily act to maintain disc height. Additionally, lumbar osteotomy destabilizes axial rotational stability at the osteotomy site, potentially further increasing mechanical demand on posterior instrumentation. Clinical studies are needed to assess the impact of this treatment strategy.

Download full-text PDF

Source
http://dx.doi.org/10.1007/s00586-017-5222-4DOI Listing

Publication Analysis

Top Keywords

interbody fusion
16
anterior lumbar
12
lumbar interbody
12
reduced motion
12
pedicle subtraction
8
subtraction osteotomy
8
pso performed
8
lumbar osteotomy
8
anterior interbody
8
bps bps + cc
8

Similar Publications

Study Design: Retrospective cohort study.

Objective: Frailty is defined as a state of minimal "physiologic reserve." The modified 5 factor frailty index (mFI-5) is a recently proposed metric for assessing frailty and has been previously studied as a predictor of morbidity and mortality.

View Article and Find Full Text PDF

Backgrounds: Postoperative inflammatory parameters are important markers of surgical site infection. Some authors have reported that spine surgery with instrumentation elevates CRP levels more than that without instrumentation does. However, those studies compared early postoperative inflammatory markers with or without instrumentation in different patients, although CRP levels vary widely among patients.

View Article and Find Full Text PDF

Purpose: This study aimed to compare the incidence of radiological adjacent segment disease (R-ASD) at L3/4 between patients with L4/5 degenerative spondylolisthesis (DS) who underwent L4/5 posterior lumbar interbody fusion (PLIF) and those who underwent microscopic bilateral decompression via a unilateral approach (MBDU) at L4/5. Our ultimate goal was to distinguish the course of natural lumbar degeneration from fusion-related degeneration while eliminating L4/5 decompression as a confounder.

Methods: Ninety patients with L4/5 DS who underwent L4/5 PLIF (n = 53) or MBDU (n = 37) and were followed for at least 5 years were retrospectively analyzed.

View Article and Find Full Text PDF

Objectives: The advent of O-arm navigation optimized the oblique lumbar interbody fusion (OLIF) procedure, allowing the operator to simultaneously perform OLIF and percutaneous posterior pedicle screw implantation without patient position change, thus improving the fluency and accuracy of the OLIF procedure (called as OLIF360). Nevertheless, a consensus regarding its suitability for patients with severe spinal stenosis remains elusive. This study aims to investigate the clinical efficacy of OLIF360 and its imaging changes in severe lumbar spinal stenosis cases.

View Article and Find Full Text PDF

Background: Currently, some novel rods with lower elastic modulus have the potential as alternatives to traditional titanium alloy rods in lumbar fusion. However, how the elastic modulus of the rod (rod-E) influences the biomechanical performance of lumbar interbody fusion remains unclear. This study aimed to explore the quantitative relationships between rod-E and the biomechanical performance of transforaminal lumbar interbody fusion (TLIF).

View Article and Find Full Text PDF

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!