Background: Preclinical evaluation of new chemical entities (NCEs) designed to be administered by inhalation route requires lung administration to rodents, especially in the discovery phase. Different administration methods have been used until now, but more efforts are required to obtain controlled and reproducible lung deposition when only small amounts of neat powder material are available.

Methods: The PreciseInhale platform used in the present study enables well-controlled powder aerosol exposures with only small amounts of micronized neat material, providing data on inhalation pharmacokinetic (PK) of NCEs at a very early stage. The DustGun aerosol technology uses compressed air to generate a respirable aerosol from milligram-amounts of powder that is delivered to one animal at a time. The new methodology was used to investigate the inhalation PK and lung retention in the rat of the novel Chiesi PDE4 inhibitor CHF6001 in three exposure models of the PreciseInhale platform: nose-only, intratracheally intubated rat, and the isolated, ventilated, and perfused rat lung. Results were compared with data from two other pulmonary delivery systems commonly used in preclinical studies: liquid instillation and powder insufflation.

Results: Administration of micronized CHF6001 using the PreciseInhale system yielded lung exposures in the same range as the other tested devices, but the reproducibility in lung deposition was improved. The initial amount of CHF6001 in lungs at the first sampling time point was close to the predetermined target dose. Tracheal deposition with PreciseInhale (0.36 ± 0.22 μg) was significantly less than with other tested delivery systems: PennCentury (23.7 ± 3.2 μg) and Airjet (25.6 ± 7.2 μg).

Conclusions: The PreciseInhale platform enabled the administration of CHF6001 powder with good accuracy and reproducibility, with low tracheal deposition. The new platform can be used at an early discovery stage to obtain inhalatory PK data for respirable aerosols of neat NCE powder without excipients and with minimal use of dry powder formulation work.

Download full-text PDF

Source
http://dx.doi.org/10.1089/jamp.2017.1369DOI Listing

Publication Analysis

Top Keywords

preciseinhale platform
12
pde4 inhibitor
8
inhibitor chf6001
8
lung deposition
8
small amounts
8
delivery systems
8
tracheal deposition
8
powder
7
preciseinhale
6
lung
6

Similar Publications

Regional lung targeting with a fluticasone/salmeterol aerosol using a bolus breath hold method of the PreciseInhale® system: A first evaluation in humans.

Eur J Pharm Sci

May 2024

Clinical Trial Consultants AB, Dag Hammarskjölds väg 10B, Uppsala SE-752 37, Sweden; Department of Global Public Health, Karolinska Institutet, SE-171 77 Stockholm, Sweden; Department of Women's and Children's Health, Uppsala University, Stockholm, Sweden.

Background: In development of inhaled drugs- and formulations the measured concentration in the systemic circulation is often used as a surrogate for local dosimetry in the lungs. To further elucidate regional differences in the fate of drugs in the lungs, different aerodynamic sizes of aerosols have been used to target major airway regions. An alternative approach to achieve regional targeting of aerosols, is to use a defined aerosol bolus together with a bolus breath hold strategy.

View Article and Find Full Text PDF

Many substances used in inhalation research are water soluble and can be administered as nebulized solutions. Typical examples are therapeutic, small-molecular agents, or macromolecules. Another category is a number of water-soluble agents used for airway diagnostics or disease modeling.

View Article and Find Full Text PDF

Background: Preclinical evaluation of new chemical entities (NCEs) designed to be administered by inhalation route requires lung administration to rodents, especially in the discovery phase. Different administration methods have been used until now, but more efforts are required to obtain controlled and reproducible lung deposition when only small amounts of neat powder material are available.

Methods: The PreciseInhale platform used in the present study enables well-controlled powder aerosol exposures with only small amounts of micronized neat material, providing data on inhalation pharmacokinetic (PK) of NCEs at a very early stage.

View Article and Find Full Text PDF

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!