Background: Snake-like robot is an emerging form of serial-link manipulator with the morphologic design of biological snakes. The redundant robot can be used to assist medical experts in accessing internal organs with minimal or no invasion. Several snake-like robotic designs have been proposed for minimal invasive surgery, however, the few that were developed are yet to be fully explored for clinical procedures. This is due to lack of capability for full-fledged spatial navigation. In rare cases where such snake-like designs are spatially flexible, there exists no inverse kinematics (IK) solution with both precise control and fast response.

Methods: In this study, we proposed a non-iterative geometric method for solving IK of lead-module of a snake-like robot designed for therapy or ablation of abdominal tumors. The proposed method is aimed at providing accurate and fast IK solution for given target points in the robot's workspace. n-1 virtual points (VPs) were geometrically computed and set as coordinates of intermediary joints in an n-link module. Suitable joint angles that can place the end-effector at given target points were then computed by vectorizing coordinates of the VPs, in addition to coordinates of the base point, target point, and tip of the first link in its default pose. The proposed method is applied to solve IK of two-link and redundant four-link modules.

Results: Both two-link and four-link modules were simulated with Robotics Toolbox in Matlab 8.3 (R2014a). Implementation result shows that the proposed method can solve IK of the spatially flexible robot with minimal error values. Furthermore, analyses of results from both modules show that the geometric method can reach 99.21 and 88.61% of points in their workspaces, respectively, with an error threshold of 1 mm. The proposed method is non-iterative and has a maximum execution time of 0.009 s.

Conclusions: This paper focuses on solving IK problem of a spatially flexible robot which is part of a developmental project for abdominal surgery through minimal invasion or natural orifices. The study showed that the proposed geometric method can resolve IK of the snake-like robot with negligible error offset. Evaluation against well-known methods shows that the proposed method can reach several points in the robot's workspace with high accuracy and shorter computational time, simultaneously.

Download full-text PDF

Source
http://www.ncbi.nlm.nih.gov/pmc/articles/PMC5540246PMC
http://dx.doi.org/10.1186/s12938-017-0383-2DOI Listing

Publication Analysis

Top Keywords

proposed method
20
snake-like robot
12
spatially flexible
12
geometric method
12
non-iterative geometric
8
inverse kinematics
8
minimal invasion
8
proposed
8
study proposed
8
method
8

Similar Publications

Background: Machine learning models can reduce the burden on doctors by converting medical records into International Classification of Diseases (ICD) codes in real time, thereby enhancing the efficiency of diagnosis and treatment. However, it faces challenges such as small datasets, diverse writing styles, unstructured records, and the need for semimanual preprocessing. Existing approaches, such as naive Bayes, Word2Vec, and convolutional neural networks, have limitations in handling missing values and understanding the context of medical texts, leading to a high error rate.

View Article and Find Full Text PDF

Background: The increasing use of social media to share lived and living experiences of substance use presents a unique opportunity to obtain information on side effects, use patterns, and opinions on novel psychoactive substances. However, due to the large volume of data, obtaining useful insights through natural language processing technologies such as large language models is challenging.

Objective: This paper aims to develop a retrieval-augmented generation (RAG) architecture for medical question answering pertaining to clinicians' queries on emerging issues associated with health-related topics, using user-generated medical information on social media.

View Article and Find Full Text PDF

Predicting protein-protein interactions (PPIs) is crucial for advancing drug discovery. Despite the proposal of numerous advanced computational methods, these approaches often suffer from poor usability for biologists and lack generalization. In this study, we designed a deep learning model based on a coattention mechanism that was capable of both PPI and site prediction and used this model as the foundation for PPI-CoAttNet, a user-friendly, multifunctional web server for PPI prediction.

View Article and Find Full Text PDF

Objective: Radiographic findings to identify central venous catheter misplacement in the arteries, which can cause lethal complications, have not been fully evaluated, and its training is difficult because it is rare. The purpose of this study is to clarify radiographic findings for differentiating central venous and misplaced arterial lines using virtual chest radiographs and elucidate their usefulness in training radiologists.

Methods: This retrospective study included 150 patients (mean age, 67 [SD, ±12] years; 97 men) who underwent colon cancer surgery between January 2018 and December 2020.

View Article and Find Full Text PDF

Objective: Patient characteristics, iodine injection, and scanning parameters can impact the quality and consistency of contrast enhancement of hepatic parenchyma in CT imaging. Improving the consistency and adequacy of contrast enhancement can enhance diagnostic accuracy and reduce clinical practice variability, with added positive implications for safety and cost-effectiveness in the use of contrast medium. We developed a clinical tool that uses patient attributes (height, weight, sex, age) to predict hepatic enhancement and suggest alternative injection/scanning parameters to optimize the procedure.

View Article and Find Full Text PDF

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!