Objective: To compare the initial biomechanical properties of olecranon osteotomies stabilized with intramedullary pins and a Nitinol bone staple to osteotomies stabilized with pin and tension band wire fixation.

Study Design: Ex vivo mechanical evaluation on cadaveric bones.

Material And Methods: Ten pairs of cadaveric forelimbs from skeletally mature Greyhounds with an olecranon osteotomy stabilized with either a pin and Nitinol bone staple or a pin and tension band wire. A single load to failure was applied to each specimen through the triceps tendon. Biomechanical properties were compared based on stiffness, yield load, and maximum load to failure and load at 2 mm of axial displacement.

Results: Specimens stabilized with the bone staple were biomechanically superior in all the variables tested. There was significantly greater stiffness (118.0 ± 25.9 N/mm versus 70.1 ± 40.4 N/mm; p = 0.005), yield load (319.0 ± 99.8 N versus 238.0 ± 42.5 N; p = 0.03), maximum load sustained (385.0 ± 99.2 N versus 287.0 ± 37.4 N; p = 0.009), and load at 2 mm of axial displacement (218.0 ± 51.5 N versus 138.0 ± 48.7 N; p = 0.002) in specimens stabilized with pins and a Nitinol bone staple than specimens stabilized with pin and tension band wire fixation.

Clinical Significance: The pin and Nitinol bone staple construct provides a biomechanically superior alternative to pin and tension band wire fixation for stabilization of olecranon osteotomies, and its use warrants further clinical investigation.

Download full-text PDF

Source
http://dx.doi.org/10.3415/VCOT-17-02-0025DOI Listing

Publication Analysis

Top Keywords

bone staple
24
nitinol bone
20
pin tension
20
tension band
20
band wire
20
pin nitinol
12
olecranon osteotomies
12
stabilized pin
12
specimens stabilized
12
pin
8

Similar Publications

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!