Objectives: Accurate sizing of the left atrial appendage (LAA) is essential when performing percutaneous LAA closure. This study aimed to compare different LAA imaging modalities and sizing methods in order to obtain successful LAA closure.

Background: Percutaneous LAA closure is an increasingly used treatment strategy to prevent stroke in patients with atrial fibrillation. LAA sizing has typically been done by 2D-transoesophageal echocardiography (TEE).

Methods: Patients who had a preprocedural TEE and preprocedural and postprocedural multislice CT (MSCT) were identified. Preprocedural measurements of LAA ostia and landing zones by 2D-TEE, MSCT and angiography were collected and analysed for those patients with successful LAA closure - i.e. with no contrast leakage at 3-month follow-up MSCT.

Results: The study population (n=67) had a mean CHADS-VASc score of 3.0 and HAS-BLED score of 2.7. Fifty-eight patients (87%) were identified to have successful LAA closure. Based on MSCT, 48 LAA sizings (83%) resulted in a correct LAA closure device size selection, whereas with 2D-TEE sizing, only 33 measurements (57%) would have resulted in a correct device size selection (p<0.01). Using adapted Bland-Altman method, MSCT-based perimeter-derived mean diameter was shown to be the best parameter to guide LAA device size selection for ‘closed-end’ devices (Amulet, WatchmanFLX), whereas the maximal diameter was the best parameter for the ‘open-end’ Watchman device.

Conclusions: Preprocedural MSCT-based LAA closure device size selection proves to be a more accurate method than conventional 2D-TEE-based sizing. Depending on the LAA closure device design, perimeter-derived mean diameter or maximal diameter could be the better sizing method.

Download full-text PDF

Source
http://www.ncbi.nlm.nih.gov/pmc/articles/PMC5515183PMC
http://dx.doi.org/10.1136/openhrt-2017-000627DOI Listing

Publication Analysis

Top Keywords

laa closure
20
successful laa
12
laa
11
imaging modalities
8
left atrial
8
atrial appendage
8
percutaneous laa
8
device size
8
size selection
8
closure
6

Similar Publications

(1) Background: Cerebral magnetic resonance imaging has reported new cerebral ischemic lesions after left atrial appendage (LAA) closure in about one- third of patients. Stroke occurs predominantly periprocedurally. This study evaluated the characteristics of embolized debris captured by the SENTINEL cerebral embolic protection system in patients undergoing LAA closure; (2) Methods: Sixty filters of 30 consecutive patients undergoing LAA closure with the WATCHMAN FLX device were collected and captured debris was analyzed by histopathology and histomorphometry.

View Article and Find Full Text PDF

Using transthoracic echocardiography (TTE) and transesophageal echocardiography (TEE) to investigate the occurrence and related causes of iatrogenic atrial septal defect (iASD) after catheter ablation combined with left atrial appendage closure (LAAC) for atrial fibrillation (AF) and its impact on the right heart system. We retrospectively analyzed 330 patients that underwent combined procedure of catheter ablation for AF and LAAC at General Hospital of Northern Theater Command from January 2018 to March 2022. These patients were divided into iASD group and non-iASD group according to whether there was persistent iASD shown on TEE at 3 months after procedure.

View Article and Find Full Text PDF

"Perpsectives on Mini-invasive left atrial appendage closure : preliminary clinical outcomes and cost data.".

Can J Cardiol

January 2025

Cardiology Department, CHU Clermont-Ferrand, F-63000 Clermont-Ferrand, France; Université Clermont Auvergne, CHU Clermont-Ferrand, CNRS, SIGMA Clermont, Institut Pascal, F-63000 Clermont-Ferrand, France.

View Article and Find Full Text PDF

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!