Purpose: The aim of this study is to compare two different methods of analysis of preoperative reference marking for toric intraocular lens (IOL) after marking with an electronic marker.

Setting/venue: Cataract and IOL Implantation Service, Shroff Eye Centre, New Delhi, India.

Patients And Methods: Fifty-two eyes of thirty patients planned for toric IOL implantation were included in the study. All patients had preoperative marking performed with an electronic preoperative two-step toric IOL reference marker (ASICO AE-2929). Reference marks were placed at 3-and 9-o'clock positions. Marks were analyzed with two systems. First, slit-lamp photographs taken and analyzed using Adobe Photoshop (version 7.0). Second, Tracey iTrace Visual Function Analyzer (version 5.1.1) was used for capturing corneal topograph examination and position of marks noted. Amount of alignment error was calculated.

Results: Mean absolute rotation error was 2.38 ± 1.78° by Photoshop and 2.87 ± 2.03° by iTrace which was not statistically significant ( = 0.215). Nearly 72.7% of eyes by Photoshop and 61.4% by iTrace had rotation error ≤3° ( = 0.359); and 90.9% of eyes by Photoshop and 81.8% by iTrace had rotation error ≤5° ( = 0.344). No significant difference in absolute amount of rotation between eyes when analyzed by either method.

Conclusions: Difference in reference mark positions when analyzed by two systems suggests the presence of varying cyclotorsion at different points of time. Both analysis methods showed an approximately 3° of alignment error, which could contribute to 10% loss of astigmatic correction of toric IOL. This can be further compounded by intra-operative marking errors and final placement of IOL in the bag.

Download full-text PDF

Source
http://www.ncbi.nlm.nih.gov/pmc/articles/PMC5516471PMC
http://dx.doi.org/10.4103/ojo.OJO_163_2013DOI Listing

Publication Analysis

Top Keywords

toric iol
12
rotation error
12
electronic preoperative
8
preoperative reference
8
reference marker
8
toric intraocular
8
intraocular lens
8
methods analysis
8
adobe photoshop
8
iol implantation
8

Similar Publications

Background: This study aims to comprehensively evaluate the predictive accuracy of six widely used toric intraocular lens (IOL) calculators in eyes undergoing cataract surgery with toric IOL implantation.

Methods: This retrospective study reviewed 53 eyes of 53 patients that underwent cataract extraction with toric IOL implantation using Zeiss 709 M. Six toric IOL calculators were evaluated: Barrett toric calculator (with predicted PCA, measured PCA, and TK), Kane formula (predicted PCA), and EVO 2.

View Article and Find Full Text PDF

Objective: Toric intraocular lens implantation has been used to correct corneal astigmatism during cataract surgery. The study aimed to compare the visual outcomes between manual vs markerless toric intraocular lens implantation in astigmatic correction.

Methods: The medical records of patients at American Eye Center who underwent phacoemulsification by multiple surgeons with insertion of monofocal or multifocal toric lenses via manual marking and markerless method from 2010-2019 were reviewed.

View Article and Find Full Text PDF

Background: The Eyhance Toric intraocular lens (IOL) builds upon the Tecnis Toric platform, initially associated with considerable post-operative rotational instability. Version 2, the Eyhance Toric IOL has been modified to enhance rotational stability. This study evaluates the post-operative rotational stability of the Eyhance Toric IOL compared to the Clareon Toric IOL, recognized for its stable performance.

View Article and Find Full Text PDF

Background: To determine whether accounting for posterior corneal surgically induced astigmatism (SIA) would improve toric intraocular lens power calculation prediction error.

Methods: A total of 189 eyes of 148 patients undergoing routine cataract surgery were included in the study. Standard and posterior keratometry were measured pre- and postoperatively.

View Article and Find Full Text PDF

Purpose: To evaluate the accuracy of toric intraocular lens (IOL) axis prediction between two preoperative measurement devices: the optical biometry (IOLMaster 500 or 700) and the dual Scheimpflug topography (Galilei G4).

Methods: Medical records of 64 eyes from 44 patients who underwent phacoemulsification and posterior chamber toric IOL (Zeiss AT TORBI 709M) implantation between July 2017 and January 2022 were reviewed. All patients underwent preoperative evaluation by optical biometry (IOLMaster 500 or IOLMaster 700) and Galilei G4.

View Article and Find Full Text PDF

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!