Effect of cleaning status on accuracy and precision of oxygen flowmeters of various ages.

Vet Anaesth Analg

Department of Statistics, College of Arts and Sciences, Kansas State University, KS, USA.

Published: July 2017

Objective: To evaluate oxygen flowmeters for accuracy and precision, assess the effects of cleaning and assess conformity to the American Society for Testing Materials (ASTM) standards.

Study Design: Experimental study.

Methods: The flow of oxygen flowmeters from 31 anesthesia machines aged 1-45 years was measured before and after cleaning using a volumetric flow analyzer set at 0.5, 1.0, 2.0, 3.0, and 4.0 L minute. A general linear mixed models approach was used to assess flow accuracy and precision.

Results: Flowmeters 1 year of age delivered accurate mean oxygen flows at all settings regardless of cleaning status. Flowmeters ≥5 years of age underdelivered at flows of 3.0 and 4.0 L minute. Flowmeters ≥12 years underdelivered at flows of 2.0, 3.0 and 4.0 L minute prior to cleaning. There was no evidence of any beneficial effect of cleaning on accuracy of flowmeters 5-12 years of age (p > 0.22), but the accuracy of flowmeters ≥15 years of age was improved by cleaning (p < 0.05). Regardless of age, cleaning increased precision, decreasing flow variability by approximately 17%. Nine of 31 uncleaned flowmeters did not meet ASTM standards. After cleaning, a different set of nine flowmeters did not meet standards, including three that had met standards prior to cleaning.

Conclusions: Older flowmeters were more likely to underdeliver oxygen, especially at higher flows. Regardless of age, cleaning decreased flow variability, improving precision. However, flowmeters still may fail to meet ASTM standards, regardless of cleaning status.

Clinical Relevance: Cleaning anesthesia machine oxygen flowmeters improved precision for all tested machines and partially corrected inaccuracies in flowmeters ≥15 years old. A notable proportion of flowmeters did not meet ASTM standards. Cleaning did not ensure that they subsequently conformed to ASTM standards. We recommend annual flow output validation to identify whether flowmeters are acceptable for continued clinical use.

Download full-text PDF

Source
http://dx.doi.org/10.1016/j.vaa.2016.10.002DOI Listing

Publication Analysis

Top Keywords

flowmeters
16
oxygen flowmeters
16
astm standards
16
cleaning
13
years age
12
flowmeters meet
12
meet astm
12
standards cleaning
12
cleaning status
8
accuracy precision
8

Similar Publications

Sensors play an important role in both the continuous monitoring and intermittent analyses, which are essential for the study of wastewater treatment plant management and conducting related research. Given the significant environmental impact of the issues involved, accurate measurement of the volume of water flowing into and out of treatment plants is a key parameter for plant management, ecotoxicological studies and academic research programs. Traditionally, flow measurements have been based on calibrated weirs or venturi flumes, using water level measurements for conversion into flow, according to established relationships.

View Article and Find Full Text PDF

Zooplankton in the Barents Sea has been monitored on an annual autumn survey since the late 1980s, using vertical WP-2 and oblique Multiple Opening and Closing Net and Environmental Sensing System (MOCNESS) tows over the water column. Sampling with MOCNESS is used to describe the vertical distribution and more frequent sampling with WP-2 (~3:1) to describe the horizontal distribution. We use here a large cumulative data set of 874 MOCNESS and 2850 WP-2 stations with data on size-fractioned dry-weight biomass to compare the two zooplankton sampling gears.

View Article and Find Full Text PDF

Objective: The study aimed to investigate the reliability of a portable toe pressure (TP) photoplethysmography device (PPG) by comparing it to a stationary laser Doppler flowmeter (LD) used in the Helsinki University Hospital. The study evaluated if lower limb arterial circulation could be reliably evaluated with the portable PPG which is more affordable and mobile than the stationary LD.

Methods: TPs were measured from 102 toes of 54 patients in the vascular surgery outpatient clinic, vascular surgery ward and interventional radiology recovery ward of Helsinki University Hospital.

View Article and Find Full Text PDF

Objective: Assessment of feasibility and proof of concept study for microvascular reactivity in anesthetized cats.

Methods: 12, adult, healthy, purpose-bred cats were included. Cats were anesthetized and instrumented with a perivascular ultrasound transit time flow probe around the main pulmonary artery, a femoral arterial catheter, a central venous catheter, and a laser Doppler flowmeter with a heating unit.

View Article and Find Full Text PDF

Purpose: Bone marrow mesenchymal stem cells (BMSCs) have emerged as promising candidate for postoperative therapeutics in chronic limb-threatening ischemia (CLTI). Nevertheless, their effectiveness is limited by their low survival rate and impaired functionality in the ischemic microenvironment. To overcome these challenges, we have devised an innovative delivery approach to support the utilization of BMSCs in CLTI therapy.

View Article and Find Full Text PDF

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!