Background: 911 Emergency Medical Services (EMS) systems utilize supraglottic devices for either primary advanced airway management, or for airway rescue following failed attempts at direct laryngoscopy endotracheal intubation. There is, however, limited data on objective confirmation of supraglottic airway placement in the prehospital environment. Furthermore, the ability of EMS field providers to recognize a misplaced airway is unknown.
Methods: Retrospective review of patients who underwent airway management using the King LTS-D supraglottic airway in a large urban EMS system, between 3/1/15-9/30/2015. Subjective success was defined as documentation of successful airway placement by the EMS provider. Objective success was confirmed by review of waveform capnography, with the presence of a 4-phase waveform greater than 5mmHg. Sensitivity and specificity of the field provider's assessment of success were then calculated.
Results: A total of 344 supraglottic airway attempts were reviewed. No patients met obvious death criteria. 269 attempts (85.1%) met criteria for both subjective and objective success. 19 attempts (5.6%) were recognized failures by the EMS provider. 47 (13.8%) airways were misplaced but unrecognized by the EMS provider. 4 attempts (1.2%) were correctly placed but misidentified as failures, leading to the unnecessary removal and replacement of the airway. Sensitivity of the provider's assessment was 98.5%; specificity was 28.7%.
Conclusion: The use of supraglottic airway devices results in unrecognized failed placement. Appropriate utilization and review of waveform capnography may remedy a potential blind-spot in patient safety, and systemic monitoring/feedback processes may therefore be used to prevent unrecognized misplaced airways.
Download full-text PDF |
Source |
---|---|
http://dx.doi.org/10.1016/j.resuscitation.2017.07.019 | DOI Listing |
Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!