Purpose: The purpose of this article is to present the first imaging experiments to demonstrate the functional equivalence between a conventional rotational gantry and a fixed-beam imaging geometry, and the feasibility of an iterative image-reconstruction technique under gravitational deformation.
Methods And Materials: Experiments were performed using an Elekta Axesse with Agility MLC and XVI, a custom-built rotating phantom stage, a Catphan QA phantom, and a porcine heart. For the imaging equivalence, a conventional cone beam computed tomography (CBCT) of the Catphan was acquired, as well as a set of 660 x-ray projections with a static gantry and rotating Catphan. Both datasets were reconstructed with the Feldkamp-Davis-Kress (FDK) algorithm, and the resultant volumetric images were compared using standard metrics. For imaging under gravitational deformation, a conventional CBCT of the Catphan and a set of 660 x-ray projections with a static gantry and rotating Catphan were also acquired with a porcine heart. The conventional CBCT was reconstructed using FDK. The projections that were acquired with the heart rotating were sorted into angular bins and reconstructed with prior image constrained compressed sensing using a deformation-blurred FDK prior. Deformation was quantified with B-spline transformation-based deformable image registration.
Results: For imaging equivalence, the difference between the two Catphan images was consistent with Poisson noise. For imaging under gravitational deformation, the conventional CBCT porcine heart image (ground truth at 0 degrees) matched the static gantry, rotating heart reconstruction with a mean magnitude of <3 mm and maximum magnitude of <5 mm of the deformation vector field. The mean deformation of the rotating heart was 3.0 to 8.9 mm, up to 16.1 mm maximum deformation. Deformation was mainly observed in the direction of gravity.
Conclusions: We have demonstrated imaging equivalence in cone beam CT reconstructions between rigid phantom images acquired with a conventional rotating gantry and with a fixed-gantry and rotating phantom. We have presented a method for image reconstruction under a fixed-beam imaging geometry using a deformable phantom.
Download full-text PDF |
Source |
---|---|
http://www.ncbi.nlm.nih.gov/pmc/articles/PMC5514241 | PMC |
http://dx.doi.org/10.1016/j.adro.2016.10.004 | DOI Listing |
J Appl Clin Med Phys
January 2025
Department of Radiotherapy, University Medical Center Utrecht, Utrecht, Netherlands.
Introduction: This paper describes a method to improve gantry-dependent beam steering for Elekta traveling wave linear accelerators by applying the measured and filtered beam servo corrections to the existing lookup table (LUT). Beam steering has a direct influence on the treatment accuracy by affecting the beam symmetry and position. The presented method provides an improved LUT with respect to the default Elekta method to reduce treatment delivery interruptions.
View Article and Find Full Text PDFMed Phys
December 2024
The University of Texas MD Anderson Cancer Center UTHealth Houston Graduate School of Biomedical Sciences, Houston, Texas, USA.
Background: Automation in radiotherapy presents a promising solution to the increasing cancer burden and workforce shortages. However, existing automated methods for breast radiotherapy lack a comprehensive, end-to-end solution that meets varying standards of care.
Purpose: This study aims to develop a complete portfolio of automated radiotherapy treatment planning for intact breasts, tailored to individual patient factors, clinical approaches, and available resources.
Front Oncol
December 2024
Medical Physics and Biomedical Engineering Lab (MPBEL), Yonsei University College of Medicine, Seoul, Republic of Korea.
Purpose: This study presents novel quality assurance (QA) approach for volumetric modulated arc therapy (VMAT) that leverages frame-by-frame electronic portal imaging device (EPID) images integrated into Mobius3D for accurate three-dimensional dose calculations.
Methods: Sequential EPID images for VMAT plans were acquired every 0.4-second by iView system and processed through iterative deconvolution to mitigate blurring from photon scattering.
Radiology
December 2024
From the Centro Cardiologico Monzino, IRCCS, Via C. Belgioioso 173, Milan, Italy (D.A., S.M., D.T., E.C., G.L., S. Galli, G.M., L.G., G.T., S.T., S. Gili, P.M., P.O., V.M., D.M., M.S., C.G., E.M., A.B., M.E.M., A.A., A F., G.P., A.L.B.); Department of Biomedical and Clinical Sciences, University of Milan, Milan, Italy (D.A.); IRCCS Ospedale Galeazzi Sant'Ambrogio, Milan, Italy (E.C., G.M., L.G., V.M., D.M., M.S., E.G., P.P., E.M., A.L.B.); Cardiovascular Center Aalst, OLV Hospital, Aalst, Belgium (J.S., M.B., E.G., P.P., K.S., T.M., C.C.); Department of Medicine, Division of Cardiology, Showa University School of Medicine, Tokyo, Japan (K.S., T.M.); Department of Biomedical, Surgical and Dental Sciences, University of Milan, Milan, Italy (G.P.); and Department of Radiology, Azienda Ospedaliera di Padova, Padua, Italy (L.Z.).
Background The detection of in-stent restenosis (ISR) with coronary CT angiography (CCTA) is challenging, but CT perfusion (CTP) has demonstrated improved diagnostic accuracy over CCTA in patients with stents. However, there are limited data on the performance of dynamic CTP, which allows noninvasive adjudication of regional myocardial blood flow. Purpose To compare the diagnostic performance of regadenoson-stress dynamic CTP with that of CCTA, using fractional flow reserve (FFR) and the index of microvascular resistance (IMR) as reference standards for epicardial coronary circulation and coronary microcirculation, respectively.
View Article and Find Full Text PDFArXiv
November 2024
Department of Radiation Oncology, University of Kansas Medical Center, USA.
Objective: Proton spot-scanning arc therapy (ARC) is an emerging modality that can improve the high-dose conformity to targets compared with standard intensity-modulated proton therapy (IMPT). However, the efficient treatment delivery of ARC is challenging due to the required frequent energy changes during the continuous gantry rotation. This work proposes a novel method that delivers a multiple IMPT (multi-IMPT) plan that is equivalent to ARC in terms of biologically effective dose (BED).
View Article and Find Full Text PDFEnter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!