Introduction: This study evaluated the cyclic fatigue resistance (CFR) of Reciproc (R25 and R40; VDW, Munich, Germany) and Reciproc Blue (R25 and R40, VDW) instruments used in an artificial S-shaped canal.
Methods: A total of 80 files were tested in an S-shaped canal (n = 20 for each file, Reciproc R25 and R40 and Reciproc Blue R25 and R40). This study compared Reciproc R25 with Reciproc Blue R25 files and Reciproc R40 with Reciproc Blue R40 files. All files were rotated in an S-shaped artificial canal until fracture. CFR was determined by recording the time to fracture in the artificial canal. The length of each fractured fragment was measured in millimeters. An independent sample t test was used to analyze the data.
Results: Between the R25 files, Reciproc Blue instruments showed significantly greater CFR than the Reciproc files in the apical and coronal curves (P < .05). Between the R40 files, Reciproc Blue instruments exhibited greater CFR in the apical and coronal curves (P < .05). There was no difference in the fractured fragment lengths of the Reciproc Blue files compared with the Reciproc files (P > .05).
Conclusions: The Reciproc Blue R25 and R40 files showed greater CFR than the Reciproc R25 and R40 files in an S-shaped canal.
Download full-text PDF |
Source |
---|---|
http://dx.doi.org/10.1016/j.joen.2017.04.009 | DOI Listing |
J Endod
January 2025
Faculdade São Leopoldo Mandic, Instituto de Pesquisa São Leopoldo Mandic, Área de Endodontia, Campinas, São Paulo, Brasil.
Introduction: This study evaluated the shaping ability of the single-file XP-endo Shaper (XP-S) system, employing an extended working time, and the Reciproc Blue #40/.06 (REC) instrument during the preparation of oval-shaped canals.
Methods: Eleven pairs of contralateral premolars with a single canal were divided into two groups (n = 11) according to the canal preparation protocol: XP-S (with an additional 45 seconds of instrumentation) and REC.
Beilstein J Nanotechnol
January 2025
Department for Chemistry and Physics of Materials, University of Salzburg, Jakob-Haringer-Str. 2a, 5020 Salzburg, Austria.
The scales of the gold-dust weevil are green because of three-dimensional diamond-type chitin-air photonic crystals with an average periodicity of about 430 nm and a chitin fill fraction of about 0.44. A single scale usually contains one to three crystallites with different lattice orientations.
View Article and Find Full Text PDFJ Endod
December 2024
Department of Endodontics, Fluminense Federal University, Niteroi, Brazil.
Introduction: This study aimed to evaluate the mechanical properties of Reciproc, Reciproc Blue, One RECI, and R-Motion, specifically focusing on their flexibility and buckling resistance and to assess their torsional fatigue behavior under 2 different reciprocation angles (150°/30° or 70°/30°).
Methods: A total of 160 instruments (40 per brand) were tested. Flexion and buckling tests were conducted using a Universal testing machine (DL 200 MF).
J Endod
December 2024
Department of Restorative Dentistry, School of Dentistry of Ribeirão Preto, University of São Paulo (USP), Ribeirão Preto, São Paulo, Brazil. Electronic address:
Introduction: This study aimed to evaluate, using microcomputed tomography, the impact of various mechanized systems on the geometry of the root canal system and dentin thickness in mandibular first molars. The hypothesis proposed that different systems would produce significant variations in root canal system preparation.
Methods: Mesial and distal canals of mandibular molars were selected and divided into 5 groups (n = 10) based on the system used: 2Shape (25.
Braz Dent J
December 2024
Dental Specialty Center, Brazilian Military Police, Minas Gerais, Brazil.
This study assessed the impact of pre-enlarging the root canal using a size 25 K-file on the fracture incidence of three reciprocating instruments after multiple uses. After a glide path performed with a size 15 K-file, the instruments (Reciproc R25, Reciproc Blue R25, and WaveOne Gold Primary) were initially used in 936 root canals. Then, for the second and third uses (933 and 927, respectively), a pre-enlargement using a size 25 K-file at the working length was done.