Comparison of intraocular pressure measurements using Goldmann tonometer, I-care pro, Tonopen XL, and Schiotz tonometer in patients after Descemet stripping endothelial keratoplasty.

Indian J Ophthalmol

Department of Ophthalmology, Tel Aviv Sourasky Medical Center, Affiliated to the Sackler School of Medicine, Tel Aviv University, Tel Aviv, Israel.

Published: July 2017

Purpose: Post-Descemet stripping endothelial keratoplasty (DSEK) patients are prone for intraocular pressure (IOP) elevations and glaucoma. Corneal characteristics influence various IOP measuring devices in various ways. The aim of this study was to evaluate the agreement between four different IOP measuring devices: Goldmann applanation tonometer (GAT), I-care pro, Tonopen XL, and Schiotz tonometr in patients who underwent DSEK.

Methods: This was a prospective comparative study using a convenience cohort of post-DSEK patients with compact grafts. Post-DSEK patients had IOP measured using GAT, I-care Pro, Tonopen XL, and Schiotz tonometer. Measurements were compared and agreement assessed. Wilcoxon signed-rank test was used for comparison of means as variables did not show a normal distribution. Bland-Altman plots were used for assessing agreement.

Results: Thirty eyes of 24 patients were included in the study. Mean time from DSEK surgery was 25.31 ± 13.05 months. Mean IOP with GAT, I-care pro, Tonopen XL, and Schiotz tonometer was 13.99 ± 3.76, 13.92 ± 3.36, 13.31 ± 3.89, and 12.83 ± 4.07, respectively. GAT, I-care pro, and Tonopen XL had similar mean IOP measurements (P = 0.135 and P = 0.551, respectively), while Schiotz tonometry measurements were higher (P = 0.046). Bland-Altman plots show good agreement between GAT, Tonopen XL, and I-care pro. GAT and Schiotz tonometry show less agreement, with large variations in the differences of measured IOP.

Conclusions: IOP measurements in post-DSEK patients showed good agreement between GAT and either Tonopen XL or I-care pro. Schiotz tonometer has large variations in this patient group. IOP measurements and IOP difference between devices were not dependent on central corneal thickness.

Download full-text PDF

Source
http://www.ncbi.nlm.nih.gov/pmc/articles/PMC5549409PMC
http://dx.doi.org/10.4103/ijo.IJO_31_17DOI Listing

Publication Analysis

Top Keywords

i-care pro
28
pro tonopen
20
tonopen schiotz
16
schiotz tonometer
16
gat i-care
16
post-dsek patients
12
iop measurements
12
iop
9
intraocular pressure
8
stripping endothelial
8

Similar Publications

Context: Concerns about brolucizumab's (Pagenax®) association with intraocular inflammation (IOI) limit its use despite its cost-effectiveness and efficacy. This multicentric study analyzes IOI incidence across 21 tertiary eyecare centers in India since its introduction in October 2020.

Purpose: To determine the real-world incidence rate of IOI in Indian patients secondary to intravitreal brolucizumab across 21 tertiary eye care centers in India.

View Article and Find Full Text PDF
Article Synopsis
  • People's default empathy levels for distant groups are usually low.
  • Exposure to positive portrayals of a national leader can significantly boost empathy and pro-social actions toward that leader's citizens.
  • The research indicates that this effect is stronger when the leader is perceived as popular domestically, suggesting that leaders can improve inter-society relations through favorable perceptions.
View Article and Find Full Text PDF

Comparison of intraocular pressure measurements using Goldmann tonometer, I-care pro, Tonopen XL, and Schiotz tonometer in patients after Descemet stripping endothelial keratoplasty.

Indian J Ophthalmol

July 2017

Department of Ophthalmology, Tel Aviv Sourasky Medical Center, Affiliated to the Sackler School of Medicine, Tel Aviv University, Tel Aviv, Israel.

Purpose: Post-Descemet stripping endothelial keratoplasty (DSEK) patients are prone for intraocular pressure (IOP) elevations and glaucoma. Corneal characteristics influence various IOP measuring devices in various ways. The aim of this study was to evaluate the agreement between four different IOP measuring devices: Goldmann applanation tonometer (GAT), I-care pro, Tonopen XL, and Schiotz tonometr in patients who underwent DSEK.

View Article and Find Full Text PDF

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!