In external beam radiotherapy, electronic portal imaging devices (EPIDs) are frequently used for pre-treatment and for in vivo dose verification. Currently, various MR-guided radiotherapy systems are being developed and clinically implemented. Independent dosimetric verification is highly desirable. For this purpose we adapted our EPID-based dose verification system for use with the MR-Linac combination developed by Elekta in cooperation with UMC Utrecht and Philips. In this study we extended our back-projection method to cope with the presence of an extra attenuating medium between the patient and the EPID. Experiments were performed at a conventional linac, using an aluminum mock-up of the MRI scanner housing between the phantom and the EPID. For a 10 cm square field, the attenuation by the mock-up was 72%, while 16% of the remaining EPID signal resulted from scattered radiation. 58 IMRT fields were delivered to a 20 cm slab phantom with and without the mock-up. EPID reconstructed dose distributions were compared to planned dose distributions using the [Formula: see text]-evaluation method (global, 3%, 3 mm). In our adapted back-projection algorithm the averaged [Formula: see text] was [Formula: see text], while in the conventional it was [Formula: see text]. Dose profiles of several square fields reconstructed with our adapted algorithm showed excellent agreement when compared to TPS.

Download full-text PDF

Source
http://dx.doi.org/10.1088/1361-6560/aa779eDOI Listing

Publication Analysis

Top Keywords

[formula text]
12
back-projection algorithm
8
presence extra
8
extra attenuating
8
attenuating medium
8
dose verification
8
dose distributions
8
epid
5
dose
5
algorithm presence
4

Similar Publications

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!