Purpose And Hypothesis: Correct femoral component rotation at knee arthroplasty influences patellar tracking and may determine function at extremes of movement. Additionally, such malrotation may deleteriously influence flexion/extension gap geometry and soft tissue balancing kinematics. Little is known about the effect of subtle rotational change upon load transfer across the tibiofemoral articulation. Our null hypothesis was that femoral component rotation would not influence load across this joint in predictable manner.

Methods: A cadaveric study was performed to examine load transfer using the orthosensor device, respecting laxity patterns in 6° of motion, to examine load across the medial and lateral compartments across a full arc of motion. Mixed-effect modelling allowed for quantification of the effect upon load with internal and external femoral component rotation in relation to a datum in a modern single-radius cruciate-retaining primary knee design.

Results: No significant change in maximal laxity was found between different femoral rotational states. Internal rotation of the femoral component resulted in significant increase in medial compartment load transfer for knee flexion including and beyond 60°. External rotation of the femoral component within the limits studied did not influence tibiofemoral load transfer.

Conclusions: Internal rotation of the femoral component will adversely influence medial compartment load transfer and could lead to premature polyethylene wear on the medial side.

Download full-text PDF

Source
http://www.ncbi.nlm.nih.gov/pmc/articles/PMC5907629PMC
http://dx.doi.org/10.1007/s00167-017-4640-5DOI Listing

Publication Analysis

Top Keywords

femoral component
28
load transfer
20
component rotation
16
rotation femoral
12
load
9
knee arthroplasty
8
examine load
8
internal rotation
8
medial compartment
8
compartment load
8

Similar Publications

Introduction: Numerous orthopaedic procedures including dynamic hip screw plating and various osteotomies require placement of a reference guide pin or K wire to direct bone cuts or for drilling screw holes. Appropriate positioning of these wires is a critical component of surgery. Irrespective of whether one is a seasoned surgeon or an apprentice, these wires often need repositioning and readjustment.

View Article and Find Full Text PDF

Background: Dual mobility (DM) implants in total hip arthroplasty provide excellent range of motion with low dislocation rates. A complication of this design is intraprosthetic dislocation (IPD), where the polyethylene (PE) liner dissociates from the femoral head. In older designs, IPD occurred due to a small head size and late PE wear with head-capture-mechanism failure.

View Article and Find Full Text PDF

Versatility of the Anterolateral Thigh Flap for Abdominal Wall Reconstruction.

Plast Reconstr Surg Glob Open

January 2025

From the Department of Plastic Surgery and Hand Surgery, Cantonal Hospital Aarau, Aarau, Switzerland.

Background: Abdominal wall reconstruction represents an ambitious and demanding challenge. This study aimed to illustrate the versatility of the anterolateral thigh (ALT) flap in its different designs for the reconstruction of complex defects of the abdominal wall.

Methods: Charts of patients with complex abdominal wall defects who underwent a reconstruction with an ALT flap at the University Hospital of Zurich (2018-2020) were reviewed retrospectively.

View Article and Find Full Text PDF

Purpose: >The aim of this multicenter study was to analyze the potential impact of patient demographics and cementation technique towards the development of radiolucent lines (RLLs) in primary total knee arthroplasty (TKA). It was hypothesized that cementation techniques, including higher cement volume, double-layer cementation technique and hardening in full extension, reduce RLL incidence by improving stability, whereas demographic factors such as age, BMI and smoking may increase RLL risk by affecting bone quality and mechanical loading.

Methods: Altogether, 776 patients (median age: 70.

View Article and Find Full Text PDF

Background And Purpose:  Early migration of the uncemented cruciate-sacrificing rotating platform ATTUNE and Low Contact Stress (LCS) tibial components was classified as at-risk for aseptic loosening rates exceeding 6.5% at 15 years based on recent fixation-specific migration thresholds. In this secondary report of a randomized controlled trial (RCT) we aimed to evaluate whether the 5-year migration, inducible displacement, and the clinical outcome of the ATTUNE components were comparable to those of the LCS.

View Article and Find Full Text PDF

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!