AI Article Synopsis

  • The study examines how different sampling rates of a digitizer affect the performance of neutron-gamma Pulse Shape Discrimination (PSD) using a solid stilbene scintillator with a Silicon Photomultiplier.
  • It was found that lower sampling rates resulted in a worse signal-to-noise ratio, degrading PSD effectiveness, but applying filtering and digital signal processing improved the performance even at lower sampling rates like 125MSPS.
  • The research concludes that a sampling rate of 250MSPS is sufficient for effective neutron-gamma discrimination, particularly for neutron detection at energies around 125keV and above.

Article Abstract

Choosing a digitizer with an appropriate sampling rate is often a trade-off between performance and economy. The influence of sampling rates on the neutron-gamma Pulse Shape Discrimination (PSD) with a solid stilbene scintillator coupled to a Silicon Photomultiplier was investigated in this work. Sampling rates from 125MSPS to 2GSPS from a 10-bit digitizer were used to collect detector pulses produced by the interactions of a Cf-252 source. Due to the decreased signal-to-noise ratio (SNR), the PSD performance degraded with reduced sampling rates. The reason of PSD performance degradation was discussed. Then, an efficient combination of filtering and digital signal processing (DSP) was then applied to suppress the timing noise and electronic background noise. The results demonstrate an improved PSD performance especially at low sampling rates, down to 125MSPS. Using filtering and DSP, the ascribed Figure of Merit (FOM) at 125keV (± 10keV) increased from 0.95 to 1.02 at 125MSPS. At 300keV and above, all the FOMs are better than 2.00. Our study suggests that 250MSPS is a good enough sampling rate for neutron-gamma discrimination in this system in order to be sensitive to neutrons at and above ~ 125keV.

Download full-text PDF

Source
http://dx.doi.org/10.1016/j.apradiso.2017.06.036DOI Listing

Publication Analysis

Top Keywords

sampling rates
16
sampling rate
12
psd performance
12
neutron-gamma discrimination
8
coupled silicon
8
silicon photomultiplier
8
rates 125msps
8
sampling
6
study sampling
4
rate influence
4

Similar Publications

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!