Background: Electrical stimulation of the peripheral nerves has been shown to be effective in restoring sensory and motor functions in the lower and upper extremities. This neural stimulation can be applied via non-penetrating spiral nerve cuff electrodes, though minimal information has been published regarding their long-term performance for multiple years after implantation.

Methods: Since 2005, 14 human volunteers with cervical or thoracic spinal cord injuries, or upper limb amputation, were chronically implanted with a total of 50 spiral nerve cuff electrodes on 10 different nerves (mean time post-implant 6.7 ± 3.1 years). The primary outcome measures utilized in this study were muscle recruitment curves, charge thresholds, and percent overlap of recruited motor unit populations.

Results: In the eight recipients still actively involved in research studies, 44/45 of the spiral contacts were still functional. In four participants regularly studied over the course of 1 month to 10.4 years, the charge thresholds of the majority of individual contacts remained stable over time. The four participants with spiral cuffs on their femoral nerves were all able to generate sufficient moment to keep the knees locked during standing after 2-4.5 years. The dorsiflexion moment produced by all four fibular nerve cuffs in the active participants exceeded the value required to prevent foot drop, but no tibial nerve cuffs were able to meet the plantarflexion moment that occurs during push-off at a normal walking speed. The selectivity of two multi-contact spiral cuffs was examined and both were still highly selective for different motor unit populations for up to 6.3 years after implantation.

Conclusions: The spiral nerve cuffs examined remain functional in motor and sensory neuroprostheses for 2-11 years after implantation. They exhibit stable charge thresholds, clinically relevant recruitment properties, and functional muscle selectivity. Non-penetrating spiral nerve cuff electrodes appear to be a suitable option for long-term clinical use on human peripheral nerves in implanted neuroprostheses.

Download full-text PDF

Source
http://www.ncbi.nlm.nih.gov/pmc/articles/PMC5504677PMC
http://dx.doi.org/10.1186/s12984-017-0285-3DOI Listing

Publication Analysis

Top Keywords

spiral nerve
20
nerve cuff
16
cuff electrodes
16
charge thresholds
12
nerve cuffs
12
spiral
8
human peripheral
8
peripheral nerves
8
non-penetrating spiral
8
motor unit
8

Similar Publications

[Application of spiral tracheoplasty in thyroid cancer with tracheal invasion].

Zhonghua Er Bi Yan Hou Tou Jing Wai Ke Za Zhi

December 2024

Department of Otorhinolaryngology Head and Neck Surgery, the Second Xiangya Hospital, Central South University, Changsha410011, China.

To explore the application of spiral tracheoplasty in the repair of large tracheal defects after the resection of trachea invaded by thyroid cancer. A retrospective analysis was performed on the clinical data of 11 patients, including 4 males and 7 females, aged from 36 to 67 years old, with large tracheal defects after tracheal resection due to thyroid papillary carcinoma invading the trachea in the Department of Otorhinolaryngology Head and Neck Surgery, Second Xiangya Hospital, Central South University from January 2019 to January 2022. The range of tracheal defects, time of tracheal reconstruction, postoperative complications and airway were recorded, and the patients were followed up for more than 24 months.

View Article and Find Full Text PDF

Efficacy of Higher-Voltage Long-Duration Pulsed Radiofrequency for Spinal Zoster-Associated Pain: A Randomized Controlled Trial.

Pain Physician

December 2024

Department of Pain Management, West China Hospital/West China Tianfu Hospital, Sichuan University, Chengdu, Sichuan Province, China.

Background: High-voltage (65 V) long-duration pulsed radiofrequency (HL-PRF) is an effective method for managing zoster-associated pain (ZAP), though the limited efficacy of and high recurrence rates associated with the procedure present concerns.

Objectives: This study aimed to investigate the safety and effectiveness of a higher-voltage HL-PRF treatment based on the original procedure for ZAP in the spinal area.

Study Design: A prospective, randomized, controlled trial.

View Article and Find Full Text PDF

We determined the relative expression levels of the receptors , , , and and ligands , , , and with RNAseq analysis on fetal human inner ear samples, located TrkB and TrkC proteins, and quantified with in situ hybridization on histological sections between gestational weeks (GW) 9 to 19. Spiral ganglion neurons (SGNs) and satellite glia appear to be the main source of and synthesis peaks twice at GW10 and GW15-GW17. Tonotopical gradients of revert between GW8 and GW15 and follow a maturation and innervation density gradient in SGNs.

View Article and Find Full Text PDF

Changes in the position of the inferior alveolar nerve after sagittal split ramus osteotomy and its effect on postoperative neurosensory disturbances.

J Stomatol Oral Maxillofac Surg

December 2024

State Key Laboratory of Oral Diseases & National Center for Stomatology & National Clinical Research Center for Oral Diseases, West China Hospital of Stomatology, Sichuan University, Chengdu, Sichuan, China. Electronic address:

Aim Of The Study: To analyze the positional changes of the inferior alveolar nerve after sagittal split ramus osteotomy (SSRO) and assess the impact of these changes on postoperative sensory nerve disturbances.

Material And Methods: Preoperative and postoperative (one year) spiral CT scans were used to categorize the inferior alveolar nerve into the extended side (n = 30) and retrogressive side (n = 30). Measurement software was used to observe and measure positional changes of the mandibular nerve canal, followed by a comparative analysis.

View Article and Find Full Text PDF

Large-gap nerve defects require nerve guide conduits (NGCs) for complete regeneration and muscle innervation. Many NGCs have been developed using various scaffold designs and tissue engineering strategies to promote axon regeneration. Still, most are tubular with inadequate pore sizes and lack surface cues for nutrient transport, cell attachment, and tissue infiltration.

View Article and Find Full Text PDF

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!