Background: Antibodies directed against dsDNA are a highly specific diagnostic marker for the presence of systemic lupus erythematosus and of particular importance in its diagnosis. To assess anti-dsDNA antibodies, the Crithidia luciliae-based indirect immunofluorescence test (CLIFT) is one of the assays considered to be the best choice. To overcome the drawback of subjective result interpretation that inheres indirect immunofluorescence assays in general, automated systems have been introduced into the market during the last years. Among these systems is the EUROPattern Suite, an advanced automated fluorescence microscope equipped with different software packages, capable of automated pattern interpretation and result suggestion for ANA, ANCA and CLIFT analysis.
Methods: We analyzed the performance of the EUROPattern Suite with its automated fluorescence interpretation for CLIFT in a routine setting, reflecting the everyday life of a diagnostic laboratory. Three hundred and twelve consecutive samples were collected, sent to the Central Diagnostic Laboratory of the Maastricht University Medical Centre with a request for anti-dsDNA analysis over a period of 7 months.
Results: Agreement between EUROPattern assay analysis and the visual read was 93.3%. Sensitivity and specificity were 94.1% and 93.2%, respectively. The EUROPattern Suite performed reliably and greatly supported result interpretation.
Conclusions: Automated image acquisition is readily performed and automated image classification gives a reliable recommendation for assay evaluation to the operator. The EUROPattern Suite optimizes workflow and contributes to standardization between different operators or laboratories.
Download full-text PDF |
Source |
---|---|
http://dx.doi.org/10.1515/cclm-2017-0326 | DOI Listing |
Clin Chim Acta
November 2019
Department of Clinical Laboratory, Tongji Hospital of Tongji Medical College, Huazhong University of Science and Technology, Wuhan, Hubei, China.
Background: Anti-dsDNA antibody is a specific antibody in systemic lupus erythematosus (SLE). Indirect immunofluorescence test (IIFT) is a highly specific method in detecting anti-dsDNA antibody. The application of automated system has gained better consistency than manual operation.
View Article and Find Full Text PDFAnn Clin Lab Sci
January 2019
Department of Laboratory Medicine, Chungnam National University Hospital, Daejeon, Republic of Korea
Background: The detection of antinuclear antibodies (ANAs) is crucial for the diagnosis of systemic autoimmune rheumatic diseases. Here, we evaluated the performance of an automated ANA image analyzer, EUROPattern Suite.
Methods: We compared the automated and manually revised interpretations of EUROPattern Suite with conventional microscopic interpretations of indirect immunofluorescence (IIF) by calculating the analytical sensitivity and specificity for ANA detection and evaluating the concordance between these approaches for ANA detection, pattern recognition, and titer determination.
Auto Immun Highlights
September 2018
Central Diagnostic Laboratory, Maastricht University Medical Center, P. Debyelaan 25, 6229 HX, Maastricht, The Netherlands.
Purpose: Indirect immunofluorescence (IIF) on the human epithelial cell-line HEp-2 (or derivatives) serves as the gold standard in antinuclear antibody (ANA) screening. IIF, and its evaluation, is a labor-intensive method, making ANA testing a major challenge for present clinical laboratories. Nowadays, several automated ANA pattern recognition systems are on the market.
View Article and Find Full Text PDFJ Clin Lab Anal
January 2019
Department of Clinical Laboratory, Peking University First Hospital, Beijing, China.
Background: The indirect immunofluorescence assay (IIFA) for the detection of antinuclear antibodies (ANA) was firstly described in 1958 and is still considered the reference method for ANA screening. Currently, an automated processing and recognition system for standardized and efficient ANA interpretation by human epithelial (HEp-2) cell-based immunofluorescence (IIF; EUROPattern Suite, Euroimmun) is available in China.
Methods: In this study, the performance of this novel system for positive/negative classification, pattern recognition (including homogenous, speckled, nucleolar, nuclear dots, cytoplasmic, and centromeres patterns) and titers evaluation was evaluated by comparing to visual interpretation.
Front Immunol
July 2019
Department of Science and Technology, Laboratory Corporation of America Holdings (LabCorp), Elon, NC, United States.
Indirect immunofluorescence (IIF) is considered by the American College of Rheumatology (ACR) and the international consensus on ANA patterns (ICAP) the gold standard for the screening of anti-nuclear antibodies (ANA). As conventional IIF is labor intensive, time-consuming, subjective, and poorly standardized, there have been ongoing efforts to improve the standardization of reagents and to develop automated platforms for assay incubation, microscopy, and evaluation. In this study, the workflow and performance characteristics of a fully automated ANA IIF system (Sprinter XL, EUROPattern Suite, IFA 40: HEp-20-10 cells) were compared to a manual approach using visual microscopy with a filter device for single-well titration and to technologist reading.
View Article and Find Full Text PDFEnter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!