Preexposure prophylaxis programs involve frequent human immunodeficiency virus (HIV) testing. We evaluated the sensitivity of 2 antigen/antibody immunoassays (Architect and Bioplex), 2 antibody-based rapid tests (Vikia-HIV-1/2 and Autotest-VIH), and 1 antigen/antibody rapid test (Alere HIV Combo) for the diagnosis of HIV infection. Among the 31 HIV-1-infected participants in the ANRS-IPERGAY trial, HIV-1 RNA was detected alone in only 2. The sensitivities of the Architect and Bioplex assays were 83% (95% confidence interval [CI], 76%-99%) and 82% (95% CI, 63%-94%), respectively. The sensitivities of the Vikia, Autotest, and Alere tests were 54% (95% CI, 34%-72%), 50% (95% CI, 31%-69%), and 78% (95% CI, 58%-91%), respectively. Antigen/antibody tests should be preferred to avoid missing cases of acute HIV infection and to decrease the related risks of viral transmission and emergence of drug resistance.

Download full-text PDF

Source
http://dx.doi.org/10.1093/infdis/jix297DOI Listing

Publication Analysis

Top Keywords

preexposure prophylaxis
8
prophylaxis programs
8
architect bioplex
8
hiv infection
8
95%
5
assessment hiv
4
hiv screening
4
tests
4
screening tests
4
tests preexposure
4