Comparison of anti-siphon devices-how do they affect CSF dynamics in supine and upright posture?

Acta Neurochir (Wien)

Product Development Group Zurich, Department of Mechanical and Process Engineering, ETH Zurich, Tannenstrasse 3, CLA G 21.1, 8092, Zürich, Switzerland.

Published: August 2017

Background: Three different types of anti-siphon devices (ASDs) have been developed to counteract siphoning-induced overdrainage in upright posture. However, it is not known how the different ASDs affect CSF dynamics under the complex pressure environment seen in clinic due to postural changes. We investigated which ASDs can avoid overdrainage in upright posture best without leading to CSF accumulation.

Methods: Three shunts each of the types Codman Hakim with SiphonGuard (flow-regulated), Miethke miniNAV with proSA (gravitational), and Medtronic Delta (membrane controlled) were tested. The shunts were compared on a novel in vitro setup that actively emulates the physiology of a shunted patient. This testing method allows determining the CSF drainage rates, resulting CSF volume, and intracranial pressure in the supine, sitting, and standing posture.

Results: The flow-regulated ASDs avoided increased drainage by closing their primary flow path when drainage exceeded 1.39 ± 0.42 mL/min. However, with intraperitoneal pressure increased in standing posture, we observed reopening of the ASD in 3 out of 18 experiment repetitions. The adjustable gravitational ASDs allow independent opening pressures in horizontal and vertical orientation, but they did not provide constant drainage in upright posture (0.37 ± 0.03 mL/min and 0.26 ± 0.03 mL/min in sitting and standing posture, respectively). Consequently, adaptation to the individual patient is critical. The membrane-controlled ASDs stopped drainage in upright posture. This eliminates the risk of overdrainage, but leads to CSF accumulation up to the volume observed without shunting when the patient is upright.

Conclusions: While all tested ASDs reduced overdrainage, their actual performance will depend on a patient's specific needs because of the large variation in the way the ASDs influence CSF dynamics: while the flow-regulated shunts provide continuous drainage in upright posture, the gravitational ASDs allow and require additional adaptation, and the membrane-controlled ASDs show robust siphon prevention by a total stop of drainage.

Download full-text PDF

Source
http://dx.doi.org/10.1007/s00701-017-3249-2DOI Listing

Publication Analysis

Top Keywords

upright posture
20
csf dynamics
12
drainage upright
12
asds
10
affect csf
8
overdrainage upright
8
sitting standing
8
standing posture
8
gravitational asds
8
asds allow
8

Similar Publications

Syncopal reactions in blood donors: Pathophysiology, clinical course, and features.

Asian J Transfus Sci

September 2022

Department of Physiology, Mahatma Gandhi Medical College, Sri Balaji Vidyapeeth (Deemed to be University), Puducherry, India.

Vasovagal syncope (VVS) in donors is a transient loss of consciousness due to short-term global cerebral hypoperfusion, which has a rapid onset and has complete spontaneous recovery. VVS may be triggered by pain, fear, anxiety, or emotional upset and loss of blood perse. It is an exaggeration of an adaptive response meant to assist in reducing the amount of bleeding/loss of blood.

View Article and Find Full Text PDF

Aphids are observed on various plant species, with most aphids feeding downward on stems. In this study, I studied the variations in feeding postures of aphids and their mechanisms. My field observations revealed that the majority of individuals from most species fed facing downward, or more precisely, towards the roots.

View Article and Find Full Text PDF

In a visual inverted pendulum balancing task avoiding impending falls gets harder as we age.

Exp Brain Res

January 2025

Ashton Graybiel Spatial Orientation Laboratory, Brandeis University, MS 033, 415 South Street, Waltham, MA, 02453, USA.

Younger adults (YA) and older adults (OA) used a joystick to stabilize an unstable visual inverted pendulum (VIP) with a fundamental frequency (.27 Hz) of half that of bipedal human sway. Their task was to keep the VIP upright and to avoid ± 60° "fall" boundaries.

View Article and Find Full Text PDF

Our ability to balance upright provides a stable platform to perform daily activities. Balance deficits associated with various clinical conditions may affect activities of daily living, highlighting the importance of quantifying standing balance in ecological environments. Although typically performed in laboratory settings, the growing availability of low-cost inertial measurement units (IMUs) allows the assessment of balance in the real world.

View Article and Find Full Text PDF

Background: Lifting is generally considered as a risk factor for low back pain. A thorough investigation of the muscle function during lifting is essential for a better assessment of the potential risk of muscle impairment and towards improvements in lifting strategy. We aimed to compare the activities of the trunk muscles between the stoop and the squat lifting tasks.

View Article and Find Full Text PDF

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!