The scanning light scattering profiler (SLSP) methodology has been developed for the full-angle quantitative evaluation of forward and backward light scattering from intraocular lenses (IOLs) using goniophotometer principles. This protocol describes the SLSP platform and how it employs a 360° rotational photodetector sensor that is scanned around an IOL sample while recording the intensity and location of scattered light as it passes through the IOL medium. The SLSP platform can be used to predict, non-clinically, the propensity for current and novel IOL designs and materials to induce light scatter. Non-clinical evaluation of light scattering properties of IOLs can significantly reduce the number of patient complaints related to unwanted glare, glistening, optical defects, poor image quality, and other phenomena associated with the unintended light scattering. Future studies should be conducted to correlate SLSP data with clinical results to help identify which measured light scatter is most problematic for patients that have undergone cataract surgery subsequent to IOL implantation.
Download full-text PDF |
Source |
---|---|
http://www.ncbi.nlm.nih.gov/pmc/articles/PMC5608261 | PMC |
http://dx.doi.org/10.3791/55421 | DOI Listing |
Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!