Skin dose mapping for non-uniform x-ray fields using a backscatter point spread function.

Proc SPIE Int Soc Opt Eng

Toshiba Stroke and Vascular Research Center, University at Buffalo, Buffalo, NY, USA.

Published: February 2017

Beam shaping devices like ROI attenuators and compensation filters modulate the intensity distribution of the x-ray beam incident on the patient. This results in a spatial variation of skin dose due to the variation of primary radiation and also a variation in backscattered radiation from the patient. To determine the backscatter component, backscatter point spread functions (PSF) are generated using EGS Monte-Carlo software. For this study, PSF's were determined by simulating a 1 mm beam incident on the lateral surface of an anthropomorphic head phantom and a 20 cm thick PMMA block phantom. The backscatter PSF's for the head phantom and PMMA phantom are curve fit with a Lorentzian function after being normalized to the primary dose intensity (PSF). PSF is convolved with the primary dose distribution to generate the scatter dose distribution, which is added to the primary to obtain the total dose distribution. The backscatter convolution technique is incorporated in the dose tracking system (DTS), which tracks skin dose during fluoroscopic procedures and provides a color map of the dose distribution on a 3D patient graphic model. A convolution technique is developed for the backscatter dose determination for the non-uniformly spaced graphic-model surface vertices. A Gafchromic film validation was performed for shaped x-ray beams generated with an ROI attenuator and with two compensation filters inserted into the field. The total dose distribution calculated by the backscatter convolution technique closely agreed with that measured with the film.

Download full-text PDF

Source
http://www.ncbi.nlm.nih.gov/pmc/articles/PMC5478926PMC
http://dx.doi.org/10.1117/12.2254257DOI Listing

Publication Analysis

Top Keywords

dose distribution
20
skin dose
12
convolution technique
12
dose
10
backscatter point
8
point spread
8
compensation filters
8
beam incident
8
head phantom
8
primary dose
8

Similar Publications

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!