Aims: The global, prospective AMPLATZER Amulet observational study documents real-world periprocedural, transoesophageal echocardiographic (TEE) and clinical outcomes from left atrial appendage occlusion (LAAO) using the AMPLATZER Amulet device. The aim of this report is to describe the periprocedural and early clinical/TEE results from this study.

Methods And Results: This multicentre prospective real-world registry included 1,088 patients (75±8.5 years, 64.5% male, CHA2DS2-VASc: 4.2±1.6, HAS-BLED: 3.3±1.1) with non-valvular atrial fibrillation; 82.8% of patients were considered to have an absolute or relative contraindication to long-term anticoagulation and 72.4% had had a previous major bleeding. Periprocedural results, clinical outcomes up to the first three months and the available TEE results from the first scheduled follow-up (one to three months post implant) are reported. Successful device implantation was achieved in 99.0% of patients. During the procedure and index hospitalisation, major adverse events occurred in 3.2% of patients. Patients were discharged on a single antiplatelet agent (23.0%), dual antiplatelets (54.3%) or an oral anticoagulant (18.9%). TEE follow-up 67±23 days post procedure in 673 patients showed adequate (<3 mm jet) occlusion of the appendage in 98.2% of patients and device thrombus in 10 patients (1.5%), as evaluated by core laboratory analysis.

Conclusions: This large real-world prospective registry of catheter-based LAAO using the AMPLATZER Amulet device reports a high implant success rate and a low periprocedural complication rate in a population with a high risk of stroke and bleeding. Transoesophageal echo data confirm good closure rates during follow-up and low rates of device-associated thrombus.

Download full-text PDF

Source
http://dx.doi.org/10.4244/EIJ-D-17-00493DOI Listing

Publication Analysis

Top Keywords

amplatzer amulet
12
left atrial
8
atrial appendage
8
appendage occlusion
8
amulet device
8
periprocedural early
8
global prospective
8
observational study
8
clinical outcomes
8
three months
8

Similar Publications

Background: The new Amplatzer Steerable Delivery Sheath is a delivery system designed to improve ease-of-use and procedural results of left atrial appendage closure (LAAC). We aimed to compare procedural results after switching our LAAC program at a tertiary care center with the Amulet device to the Steerable Delivery Sheath, with a control group of LAAC employing the standard sheath.

Methods: The first n = 32 consecutively treated patients at our site using the Amulet device with the Steerable Delivery Sheath were included in this retrospective analysis.

View Article and Find Full Text PDF

Background: The Amulet IDE trial (AMPLATZER Amulet Left Atrial Appendage Occluder [LAAO] Investigational Device Exemption [IDE] Trial) evaluated the safety and effectiveness of the Amulet occluder (Abbott) in patients with nonvalvular atrial fibrillation. The Amulet IDE trial is the largest randomized LAAO trial, comparing the Amulet occluder with the Watchman 2.5 device (Boston Scientific).

View Article and Find Full Text PDF

Background: Chronic kidney disease (CKD) and end stage renal disease (ESRD) are associated with increased risk of bleeding events, including hemorrhagic stroke, and periprocedural and gastrointestinal bleeding among patients with atrial fibrillation who are on anticoagulation. Safety of percutaneous left atrial appendage occlusion (LAAO) among this patient population has been uncertain with studies showing contradictory results.

Methods: PubMed and Google Scholar databases were queried for studies comparing outcomes among patients with and without significant CKD, and with and without ESRD who underwent LAAO device implantation.

View Article and Find Full Text PDF

Clinical outcome and intraprocedural characteristics of left atrial appendage occlusion: a comparison between single-occlusive plug-type and dual-occlusive disc-type devices.

Front Cardiovasc Med

July 2024

Deutsches Herzzentrum der Charité, Department of Cardiology, Angiology and Intensive Care Medicine, Campus Virchow Klinikum, Berlin, Germany.

Article Synopsis
  • Percutaneous interventional left atrial appendage occlusion (LAAO) is a proven method for preventing strokes in patients with atrial fibrillation (AF) and has been evaluated in a study involving 149 patients at a Berlin hospital from 2016 to 2022.
  • The study compared two types of devices used in LAAO: single-occlusive plug-type (SOPT) and dual-occlusive disc-type (DODT), noting that while DODT procedures took longer, both had successful implantation with low rates of complications and no hospital deaths.
  • After six months, findings indicated that SOPT had a higher rate of device-related thrombus formation compared
View Article and Find Full Text PDF