Purpose: The presence of metallic debris near total hip arthroplasty can have a significant impact on longitudinal patient management. Methods for magnetic resonance imaging-based quantification of metallic debris near painful total hip replacements are described and applied to cohorts of symptomatic and control subject cases.
Methods: A combination of metal artifact reduction, off-resonance mapping, off-resonance background removal, and spatial clustering methods are utilized to quantify off-resonance signatures in cases of suspected metallosis. These methods are applied to a cohort of symptomatic hip arthroplasties composed of cobalt-chromium alloys. Magnetostatic simulations and theoretical principles are used to illuminate the potential sources of the measured off-resonance effects. Reported metrics from histological tissue assays extracted during surgical revision procedures are also correlated with the proposed magnetic resonance imaging-based quantification results.
Results: The presented methods identified quantifiable metallosis signatures in more than 70% of the symptomatic and none of the control cases. Preliminary correlations of the MR data with direct histological evaluation of retrieved tissue samples indicate that the observed off-resonance effect may be related to tissue necrosis.
Conclusions: Magnetostatic simulations, theoretical principles, and preliminary histological trends suggest that disassociated cobalt is the source of the observed off-resonance signature. Magn Reson Med 79:1628-1637, 2018. © 2017 International Society for Magnetic Resonance in Medicine.
Download full-text PDF |
Source |
---|---|
http://www.ncbi.nlm.nih.gov/pmc/articles/PMC5741538 | PMC |
http://dx.doi.org/10.1002/mrm.26807 | DOI Listing |
Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!