Objective: Separately addressing the fatigue resistance (ISO 14801, evaluation of final product) and aging behavior (ISO 13356, standardized sample) of oral implants made from yttria-stabilized zirconia proved to be insufficient in verifying their long-term stability, since (1) implant processing is known to significantly influence transformation kinetics and (2) aging, up from a certain level, is liable to decrease fatigue resistance. Therefore, the aim of this investigation was to apply a new testing protocol considering environmental conditions adequately inducing aging during dynamic fatigue.

Methods: Zirconia implants were dynamically loaded (10 cycles), hydrothermally aged (85°, 60 days) or subjected to both treatments simultaneously. Subsequent, monoclinic intensity ratios (X) were obtained by locally resolved X-ray microdiffraction (μ-XRD). Transformation propagation was monitored at cross-sections by μ-Raman spectroscopy and scanning electron microscopy (SEM). Finally, implants were statically loaded to fracture. Linear regression models (fracture load) and mixed models (X) were used for statistical analyses.

Results: All treatments resulted in increased fracture load (p≤0.005), indicating the formation of transformation induced compressive stresses around surface defects during all treatment modalities. However, only hydrothermal and combinational treatment were found to increase X (p<0.001). No change in X was observed for solely dynamically loaded samples (p≥0.524). Depending on the variable observed, a monoclinic layer thickness of 1-2μm (SEM) or 6-8μm (Raman spectroscopy) was measured at surfaces exposed to water during treatments.

Significance: Hydrothermal aging was successfully induced during dynamic fatigue. Therefore, the presented setup might serve as reference protocol for ensuring pre-clinically long-term reliability of zirconia oral implants.

Download full-text PDF

Source
http://dx.doi.org/10.1016/j.dental.2017.06.002DOI Listing

Publication Analysis

Top Keywords

long-term stability
8
testing protocol
8
protocol considering
8
fatigue resistance
8
fracture load
8
stability injection-molded
4
injection-molded zirconia
4
zirconia bone-level
4
bone-level implant
4
implant testing
4

Similar Publications

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!